4
3.
Connect the Kelvin clips to the calibration standard. Make sure the clips are as close together as possible. (Fig 2)
NOTE:
Make sure the calibration standard is clean. The Kelvin clips must be connected across the diameter of the
standard. Use the abrasive pad to clean the calibration standard.
Calibration Procedures
1.
Turn the CLM100 on by depressing the green on/off button
2.
Insert the black test leads into one pair of input jacks and the red test leads into the other pair of input
jacks. (Ignore the gray marking on the banana plug.) This is very important to ensure that the meter
works correctly. If black and red test leads are mixed the unit will give a false reading. (Fig 1)
INSTRUCTIONS
(Fig 1)
(Fig 2)
NOTE:
Make sure clips are as close together as possible. Use the tips of the Kelvin clips when performing
measurements on small diameter wire.
4.
Press and hold the
“CAL”
button until all segments in the display illuminate. This completes the calibration
procedure and will ensure all measurements are as accurate as possible.
5.
Disconnect the leads from the standard.