COMBUSTION ANALYSIS

T2

T1

P2

P1

Insert flue probe into stack. 

Adjust cone so the probe tip is 

approximately at the center of 

the stack.

Insert flue probe into combustion 

input. Insert K-type temperature 

probe into input T1.

Rotate function rotary dial to 

FLUE 1 for Flue Test Screen 1 

parameters. 

STATUS

FLUE 1

AUX

FLUE 2

TEMP

Combu

stion Analy

zer

C160

CO

  C

O2

  T

EM

P

Flue 1 displays CO2, O2,  

Efficiency, Flue temperature, 

Inlet temperature. 

Rotate the function rotary dial 

to FLUE 2 for Flue Test Screen 2 

parameters. 

STATUS

FLUE 1

AUX

FLUE 2

TEMP

Combu

stion Analy

zer

C163

CO

  C

O2

  T

EM

P

Flue 2 displays CO (ppm), CO2, 

Losses calculated, Excess air %. 

Make any adjustments needed 

for proper Combustion and wait 

for analyzer to display change in 

readings. (Repeat as necessary).

BEFORE TESTING

Remove the rubber boot and red 

drain plug. Allow the water to 

drain out. Re-insert the rubber 

plug and replace boot cover.

Remove protective rubber boot. 

Remove water trap from analyzer. 

Remove particle filter from water 

trap and replace with new one. 

Reattach trap and boot.

CHECKLIST

Clean particle filter 

Water trap and probe line are 

free of condensate

All hose and thermocouple 

connections are properly 

secured

Water trap is fitted correctly

Power on and zero

Flue gas probe is sampling 

ambient FRESH air

NOTE

: If there is no hole, check with service and installation instructions of appliance for best testing 

method or check with your local authority on best practice

Summary of Contents for C164

Page 1: ...YZER BEFORE USING Register within 30 Days of Purchase to Get 10 Off Your First Year Calibration Keep up on the latest Accessories Upgrades and Videos Scan Now for Easy Online Registration or Visit www ueitest com 4 4 4 ...

Page 2: ...h built in magnets Log Print Temperature Screen Heat Exchange Integrity Test Screen LED Water Trap Indication Water Trap Window Flue Gas Inlet Connection Power button Display lines 1 through 6 Status bar line Navigation Pump T2 T1 T2 T1 P2 P1 C163 C164 Flue Test screens Auxiliary screen Status screen Menu screen USB charging port K type temperature inputs Pressure Ports C163 C164 only ...

Page 3: ...Dial to STATUS to view parameters Selected Fuel Ti Ambient Temperature ATM Atmospheric Pressure measured in mbar CAL Days remaining until calibration Press and buttons to scroll Status Bar options Time Pump Status Battery Status Fuel Type Date Change Fuel Type in STATUS SCREEN Press and buttons to the currently selected fuel Press and Hold button until arrows indicate fuel edit mode is activated T...

Page 4: ...mbustion Analyzer C163 CO C O 2 T E M P Flue 2 displays CO ppm CO2 Losses calculated Excess air Make any adjustments needed for proper Combustion and wait for analyzer to display change in readings Repeat as necessary BEFORE TESTING Remove the rubber boot and red drain plug Allow the water to drain out Re insert the rubber plug and replace boot cover Remove protective rubber boot Remove water trap...

Page 5: ...P Combustion Analyzer C161 CO C O 2 T E M P STATUS F L U E 1 AUX FLUE 2 T E M P Combustion Analyzer C163 CO C O 2 T E M P STATUS F L U E 1 AUX FLUE 2 T E M P Combustion Analyzer C161 CO C O 2 T E M P AUX displays 5 lines of editable parameters To edit line values Press button until Edit displays on the Status Bar Press and Hold until the Status Bar reads LINE 1 When performing a combustion Test Ro...

Page 6: ...Combustion Analyzer C161 CO C O 2 T E M P To print the LOG Press and buttons until PRINT is selected between two arrow icons Press to select to start printing STATUS F L U E 1 AUX FLUE 2 T E M P Combustion Analyzer C161 CO C O 2 T E M P To EXIT Press and buttons until EXIT is selected between two arrow icons Press to EXIT screen RepeatuntilreturntoMENUscreen STATUS F L U E 1 AUX FLUE 2 T E M P Com...

Page 7: ...ranty and service information UEi Test Instruments 1 800 547 5740 info ueitest com This warranty gives you specific legal rights You may also have other rights which vary from state to state Download the full Owners Manual at UEiTEST COM or SCAN THIS CODE POST TESTING BATTERIES The analyzer is fitted with 3 AA size rechargeable batter ies Observe correct polarity when replacing Always check operat...

Page 8: ... 12 Month warranty 2 Day turn around on all standard recertification services with pre approved payment Our recertification teams are factory certified and proud to provide industry leading turnaround support Our recertification centers in USA and Canada are ISO IEC 17025 2017 accredited facilities that only use EPA protocol test gases and can provide NIST certification ser vices Return freight pr...

Reviews: