After the countdown is finished and the meter is correctly set up, take
the meter into the room to be tested, make sure its pump is on and use
the probe to draw in ambient air for the meter to measure.

• Only switch on the meter in OUTDOOR, FRESH AIR
• Do not exceed the meter’s internal temperature operating range
• Do not put the meter on a hot surface
• Do not expose yourself and others to dangerous levels of CO

Printing Test Information

An optional accessory for the CO91 is an infrared thermal printer. Read
the manual supplied with the printer prior to operation. Connection to
the Co91 is detailed below.

• Infrared thermal printer - this does not require a cable to transmit 

the data but uses an infrared (IR) link similar to a TV remote 
control. The IR emitter is positioned on the top of the CO91 and 
the bottom of the printer. Ensure they are pointing at each other
and within 3 feet (1 meter), with not obstructions in the way. Data
may be lost if transmission is interrupted. Keep the CO91 pointing
at the printer until the printout has finished.

During combustion tests the CO91 will print data on request. With the
analyzer showing the “

MAIN DISPLAY

” press “           “ and current

data will be sent to the printer.

The display will show “

PRINTING

” until data transmission is complete.

After Using The Meter

After using the meter to test appliances, remove it’s probe from the flue
- IT WILL BE HOT - and let it cool. Do not put the probe in water which
will be sucked into the meter, damaging its pump and sensors.

When the meter readings return to ambient levels, switch off. The 
meter counts down from 30 before switch off with the pump running 
to self-clean its sensors.

Empty the water trap and check filter prior to storage and next use.

Tro u b l e s h o o t i n g

S p e c i f i c a t i o n s

General specifications

CO91-MAN    

P. 3

UEI CO91

503 644 8723

06-26-02

14:27

SMOKE:

NATURAL GAS

CO

PPM

00

CO MAX

PPM

TEMP.

F

----

If I See This Malfunction

Then Take This Corrective Action

• CO sensor Error (----)

• Meter was stored in a cold environment

and is not at normal working temperature

• CO sensor needs replacing

• Batteries not holding charge

• Batteries exhausted

• Meter not running on mains

• AC charger not giving correct output

adapter

• Flue blown in charger plug

• Meter does not respond to

• Water trap drain plug not installed

flue gas

• Particle filter blocked
• Probe or tubing blocked
• Pump not working or damaged with

contaminant's

• Erratic temperature readings

• Temperature plug reversed in socket

or 000 on display

• Faulty connection or break in cable or plug

Parameter

Resolution

Accuracy

Range

Temperature

1.0˚F/C

±5.0˚F (2.0˚C)

32-1112˚F

Measurement

±0.3% reading

0-600˚C

Carbon Monoxide

1 ppm

±10 ppm <100 ppm*1

0-1000 ppm

(CO) measurement

±5% of reading >100 ppm

Dimensions
Weight

2.2 lb./1 kg

Handset

7.9” (L) x 1.8” (D) x 3.5” (W) (200 mm x 45 mm x 90 mm)

Probe

11.8” (L) x 0.25” (Dia) with 7.8” long stainless steel shaft
(300 mm x 6 mm with 200 mm long stainless steel shaft)

Ambient operating

32˚ to 104˚F  (±0 to +40˚C)

range

10% to 90% RH non-condensing

Battery life

4 AA cells
>8 hours using Alkaline AA cells

AC adapter

Input: 110 V AC/220 V AC nominal

(optional)

Output: 9 V DC regulated

Temperature

Part number ATT100 (included)

probe
Carrying case

AC75 (included)

Summary of Contents for C091

Page 1: ...1 800 547 5740 Fax 503 643 6322 www ueitest com email info ueitest com CO91 INSTRUCTION MANUAL ...

Page 2: ... meter must only be used in well ventilated locations by trained and competent persons after due consideration of all the potential hazards International Symbols Controls and Indicators 1 Power Push button Press button to switch the unit ON in fresh air outside the property about to be tested This lets the meter auto calibrate its sensors properly When turned on the meter beeps four times and disp...

Page 3: ...e and temperature on its bottom line if you turn the rotary dial to TEMP and push the DOWN button until CO is displayed on the top line Use the UP DOWN ENTER keys to change settings such as time and date when the rotary dial is turned to MENU 3 PRINT BACKLIGHT button Press and release PRINT BACKLIGHT quickly to switch backlight on or off Press for more than 1 second before releasing to start the m...

Page 4: ...nsors When the meter readings return to ambient levels switch off The meter counts down from 30 before switch off with the pump running to self clean its sensors Empty the water trap and check filter prior to storage and next use Troubleshooting Specifications General specifications CO91 MAN P 3 UEI CO91 503 644 8723 06 26 02 14 27 SMOKE NATURAL GAS CO PPM 00 CO MAX PPM TEMP F If I See This Malfun...

Page 5: ...r fire as they may expose and cause injury 7 Replace battery compartment cover Electromagnetic Compatibility EMC European Council Directive 89 336 EEC requires electronic equipment not to generate electromagnetic disturbances exceeding defined levels and have adequate immunity levels for normal operation Specific standards applica ble to this meter are stated below As there are electrical products...

Page 6: ... not limited to implied warranties of merchantability and fitness for a particular purpose are limited to the express warranty UEi shall not be liable for loss of use of the instrument or other incidental or consequential damages expenses or economic loss or for any claim or claims for such damage expenses or economic loss A purchase receipt or other proof of original purchase date will be require...

Reviews: