1. Failure and error messages
Failure
encountered
Possible cause
Corrective action
Display Err -
300
Excessive cuff pressure. Cuff
inflated to maximum due to
movement of the arm or body.
The hose may be kinked or the
air connector blocked.
- Repeat measurement
- Do not move arm
- Do not talk
- Check that hose with
air connector is clear
(unobstructed)
Display Err - 1
Measurement error.
Measurement could not be
carried out due to movement
of the arm or body.
- Repeat measurement
- Do not move arm
- Do not talk
Display Err - 2
Inflation error, pressure built
up too quickly or too slowly.
Check sitting of air connector,
repeat measurement.
Fault was recognised.
- Repeat measurement
- Do not move arm
- Do not talk
Display Err - 3
Deflation rate too high or
too low.
Call the Service hotline if this
occurs repeatedly.
Irregular heartbeat, disruptive
movements, shaking,
wobbling (objects), breathing
in deeply, etc.
Repeat measurement after 3-5
minutes rest. Possible factors
(depending on the severity of
the arrhythmia) affecting the
measurement results must be
discussed with your doctor!
Unusual
measured
values
Movement or talking during
measurement, resting time
not observed, feet possibly
crossed, smoking or coffee
cunsumption.
Check conditions and repeat
measurement. Observe
instructions for use P. EN-36.
Failure
encountered
Possible cause
Corrective action
Display 0
0
Start/Stop button
accidentally pressed while
battery was being changed.
Switch unit off and then on
again with Start/Stop button.
Measurement
was interrupted
and cuff first
deflated and
then inflated
Strong pulse restlessness or
movement detected, unit
rejects measurement
If necessary, interrupt
measurement, measure again
after a break of 5 minutes.
Movement during
measurement
If necessary, interrupt
measurement, measure again
after a break of 5 minutes.
No display after
unit is switched
on
Batteries inserted
incorrectly?
Check position of batteries.
Batteries flat?
Change batteries.
Battery compartment dirty?
Clean battery compartment.
Measurement
interrupted
Batteries flat?
Change batteries.
Measured
values
extremely high
or low
Was the correct
measurement position
maintained?
Repeat measurement,
ensuring correct body posture.
Do not move your arm and do
not speak.
Was the necessary
rest observed prior to
measurement?
Repeat measurement after a
break of approx. 3-5 mins.
- Do not move arm
- Do not talk