EN-16
EN-17
Technical information
Technical information
The measured temperature is above 43.99°C and is
therefore outside the measuring range.
The measured temperature is below 32.00°C and is
therefore outside the measuring range.
The ambient temperature is above 40.00°C and is
therefore outside the permissible operating condi-
tions.
The ambient temperature is below 10.00°C and is
therefore outside the permissible operating condi-
tions.
Fever warning. Temperature higher than 37.5°C meas-
ured.
Battery flat. Please change the battery.
Device memory is full. Connect the
thermometer to the cyclotest
®
myS-
ense app to transfer the measured val-
ues to the app.
Battery low. Have a spare battery ready.
Error messages
No display after switching on. Check whether the bat-
tery is positioned correctly and the polarity (+/-) is cor-
rect. If the error persists, change the battery.
Bluetooth connection error
Sensor error. If this occurs repeatedly, please contact
the cyclotest
®
service hotline.
Measurement was cancelled prematurely. Repeat
the measurement and wait until the measurement is
compete before removing the thermometer from the
measurement point.
Calibration error. If this occurs repeatedly, please con-
tact the cyclotest
®
service hotline.
The device is in technical test mode. Switch it off and
on again after a few seconds to return to normal mode.
Technical data
Type:
Maximum thermometer, direct mode
Measuring points/body
reference points:
rectal, oral, vaginal
Temperature unit:
°C
Measuring range:
32.00°C to 43.99°C