METHOD OF AIRBORNE DETECTION 

The method of air borne detection is to go from the "gross to fine". Start off at a high sensitivity level 
and if there is too much ultrasound in the area, reduce the sensitivity, following the sound to the 
loudest point.  If necessary, place the RUBBER FOCUSING PROBE (described below) over the scanning 
module and proceed to follow the test sound to its loudest point constantly reducing the sensitivity 
while following the bar graph indicator on the display.

 

HEADPHONES 

To use, firmly plug the headphone jack into the “Phones” receptacle on the pistol 

housing and place the 

headphones over your ears. 

RUBBER FOCUSING PROBE 

The Rubber Focusing probe fills two functions:  it deflects stray ultrasounds and enhances the reception 
of weak airborne signals.  To use, simply slip it over the front of the scanning module or the contact 
module.   

NOTE:

 

To prevent damage to the module plug, always remove the module BEFORE attaching 

and/or removing the Rubber focusing Probe. 

 

USERS INSTRUCTIONS 

SCANNING MODULE 

 

Plug in to front end. 

 

Align the plug located at the rear of the module with the receptacle in the front end of the 
Pistol Housing and plug in. 

 

Start to scan the test area. 

Summary of Contents for Ultraprobe 3000

Page 1: ...1 UE SYSTEMS INC THE ULTRAPROBE 3000 MANUAL...

Page 2: ...e Contact Module Pistol Grip Housing Display Panel Sensitivity Control Dial Headphone Jack USB Port Trigger On Off Switch Battery Compartment Battery Wrist Strap Standard Accessories Headset WTG 1 War...

Page 3: ...hut Down Time User Sensitivity Defaults Store Mode Program Update Exit Scanning Module Method of Airborne Detection Headphones Rubber Focusing Probe Stethoscope Module Stethoscope Extension Kit Long R...

Page 4: ...strument it is important to review this manual before you begin inspections THERE ARE TWO MODES THAT ARE IMPORTANT TO UNDERSTAND OPERATION MODE The operation mode will be described in detail under the...

Page 5: ...of the module For placement align the plug with the corresponding receptacle in the front end of the pistol housing and plug in The Scanning Module has a piezoelectric transducer to pick up the airbo...

Page 6: ...intensity levels as dB and as a bar graph Sensitivity Level Storage location number and Battery Level Intensity levels are shown simultaneously as a numeric dB value and on a sixteen segmented bar gr...

Page 7: ...tery Remove the battery only when it cannot hold a charge and needs to be replaced If the battery is to be changed remove the cover and replace The battery is a rechargeable battery and is charged usi...

Page 8: ...ic transmitter designed to flood an area with ultrasound This Tone Generator is a WARBLE TONE GENERATOR This internationally patented transmitter sweeps through a number of ultrasonic frequencies in a...

Page 9: ...s 78 7 cm with the Stethoscope Probe OPTIONAL ACCESSORIES LONG RANGE MODULE LRM This uniquely designed module doubles the detection distance of a standard scanning module and provides a narrow 10 sens...

Page 10: ...litude of a sensed ultrasound The maximum level indicator will remain at the highest sensed intensity during an inspection until 1 A new maximum reading is detected or 2 The trigger is released and th...

Page 11: ...higher the bar graph The less sound that is received the lower the bar graph FREQUENCY This instrument is set to the peak frequency response of the transducers which is 40 kHz It is non adjustable TO...

Page 12: ...ecord will be over written TO DOWNLOAD THE INFORMATION Refer to Setup Mode Menu 01 Send Records SETUP MODE TO ENTER THE SET UP MODE Make sure the Ultraprobe is off Press Click the Sensitivity dial and...

Page 13: ...ds NOTE Hold the trigger in during any of the Set Up Mode operations or the instrument will turn off When MENU 01 Send Records is shown click the Sensitivity Control Dial and all the data will be tran...

Page 14: ...enter The Spin to the desired turn off time Click to exit MENU 04 USER SENSE SENSITIVITY DEFAULTS With experience a user will know which sensitivity level to use as the highest level This mode allows...

Page 15: ...Store operation The third click exits the Store Mode to the Main Screen The Quick mode requires one click to store the data Every time you store data the instrument will move up one record to the next...

Page 16: ...tice Download the program to your computer and follow the procedure supplied by UE Systems CAUTION Failure to follow the Program Update procedure may prevent the UP3000 from programming properly and r...

Page 17: ...ES To use firmly plug the headphone jack into the Phones receptacle on the pistol housing and place the headphones over your ears RUBBER FOCUSING PROBE The Rubber Focusing probe fills two functions it...

Page 18: ...at the thread of the rod you just unscrewed and locate a rod in the kit that has the same size thread this is the base piece Screw the Base Piece into the Stethoscope Module If all 31 78 7 cm are to b...

Page 19: ...W for a low amplitude signal usually recommended for small containers or HIGH for high amplitude In high the Warble Tone Generator will cover up to 4 000 cubic feet 121 9 cu meters of unobstructed spa...

Page 20: ...lever Move this change lever to the middle of the lock in a way that allows it to hook behind the change notch drawing 1 Set your personal combination turning the dials to the desired combination i e...

Page 21: ...21 Need further support Want information regarding products or training CONTACT UE Systems Inc 14 Hayes Street Elmsford NY 10523 USA T 914 592 1220 E info uesystems com W www uesystems com...

Reviews: