![UCAN FT-F Series Installation And Operation Manual Download Page 9](http://html1.mh-extra.com/html/ucan/ft-f-series/ft-f-series_installation-and-operation-manual_3161824009.webp)
8
After that, it becomes possible to clean the inside of the water reservoir, surface of
the oscillators, outside of the tube of the UV lamp.
[Removal method of the fan box]
: Replacement of the oscillators, the float sensors or maintenance of the water
supply solenoid valve.
Procedure
1)
Unlock the snap lock on both side of the fan box.
2)
Slide the fan box toward upper side.
Float switch
Oscillator
UV lamp
Fan
Top view
Snap lock
(similarly, the oposite side)
②
②
Water reservoir
Fan box
Fixture stay