UAS CLARCOR SMOG-HOG MSH Series Owner'S Manual Download Page 7

3

revised 03/15

SMoG-HoG

®

electronic air cleaner

support structure corners or the triangular insulators. 

Inspect filters, noting the condition (contaminant build 

up?, media separation?, bent frames?) should not 

have contaminant build up restricting airflow, repair or 

replace. Module drain sump should not have moderate 

to extreme contaminant build up. Unicell module 

support tracks should be free of contaminant build-up 

for ground contact. Module and access doors gaskets 

should be in-place and in good condition. 

  5.2  MANUAL CLEANING OF SMOG-HOG 

COMPONENTS

There are a number of methods for manual cleaning, 

certain key cleaning criteria contribute to the 

effectiveness of every method. these include the type 

of detergent, detergent strength, water temperature, 

agitation/ impingement, duration, rinse procedure and 

dry-out time. 

TYPE OF DETERGENT   

In general, the detergent used on most hydrocarbons 

(e.g., oily residues) will be alkaline in nature. It is 

extremely important that the detergent have a built-in 

buffering agent to reduce aluminum deterioration.   

DETERGENT STRENGTH          

Detergent concentration in a mixture with water varies 

with the application from 1:1 to 20:1 parts water to parts 

detergent. For any contaminant condition, the best 

course is to use a cleaning solution per the detergent 

manufacturer’s directions. More or less detergent 

may eventually be required for effective cleaning at 

reasonable detergent cost.  the recommended factory 

detergent concentration is 5% (20:1). 

WATER TEMPERATURE 

Detergents can be up to twice as effective in hot water. 

Hot water alone is very effective in softening built-up 

residue. Water temperature should be 140°F to 170°F, 

not to exceed 180°F. 

AGITATION/IMPINGEMENT           

These methods are virtually the same, with 

impingement being the most extreme form of agitation. 

any liquid movement over built-up residue will remove 

a layer, allowing detergent to work on the next layer. A 

reduction in cleaning time duration usually results. 

CLEANING CYCLE DURATION           

In most cleaning methods, adequate time should be  

allowed for the detergent to remove the contaminant 

thoroughly. reaction time will vary depending on 

detergent strength, temperature and agitation. 

Guidelines for mixing, heating and expected results are 

included on specification sheets for most detergents.  

 

!

  C A U T I O N

Never mix caustic and alkaline detergents for 

manual. Detergent mixing could cause rapid 

heat release, gel formation or some other 

undesirable condition. 

time is necessary for effective cleaning. Soaking may 

seem slower and less effective than high impingement 

and/or hot water above 180°F for cleaning action, so 

personnel should be forewarned about using excessive 

pressure or temperature to shortcut the cleaning 

process. High pressure or stream spray cleaning until 

plate edges are shiny is not effective. not only will 

penetration to the cell core not occur, but warping and 

bending of the plates may result as well. Patience and 

thoroughness of cleaning best preserve the integrity of 

the components in the long term. Soaking and gentle 

rinsing provides for best results. 

RINSE PROCEDURE           

cleaned components should be rinsed off quickly and 

thoroughly to remove remaining contaminants. Some 

detergent residue will remain if rinsing is not performed. 

the residue may contribute to voltage bleed-down when 

the unicell is placed in operation. Also, even though the 

detergent is “buffered” prolonged contact could cause 

minor corrosion. Hot water should be used for rinsing. 

DRY-OUT TIME          

unicells and filter media should be dry before the MSH 

is placed into operation. Startup of a wet system will 

cause dead short conditions to the ionizer and collector 

cell circuits. Wet unicells and filter media should be 

placed in a warm room for drying. techniques such as 

hand wiping insulators and blow drying unicells and 

filter media with compressed air will decrease drying 

time. 

  5.3 MANUAL CLEANING METHODS 

the manual cleaning method selected will depend 

on the type of contaminant, rate of deposit, facility 

limitations such as cleaning time windows (process 

downtime) and available utilities. All cleaning methods 

listed in this section are acceptable. 

SOAK TANK             

this method involves placing unicells and filter media 

in an agitated solution of hot water and detergent and 

is the most effective method. With proper detergent 

selection, this procedure will quickly remove most 

contaminants. unicells and filter media should not be 

placed in highly concentrated detergent solutions or 

allowed to soak for extended periods, (e.g., overnight), 

especially at elevated temperatures. Extended soaking 

(e.g., days) in solvent or detergent solution will

 

degrade components over time and should be avoided. 

PORTABLE PRESSURE WASHER         

a self-contained pressure washer with a spray wand 

can be an effective cleaning method, providing it is 

used with caution. Care should be taken not to expose 

the unicells to close-up and prolonged blasts of high 

pressure/temperature, causing cell plate deformity, 

requiring a replacement set of unicells and filter media. 

Summary of Contents for CLARCOR SMOG-HOG MSH Series

Page 1: ...Industrial Electrostatic Air Cleaner l SMOG HOG MSH Model OWNER S MANUAL ...

Page 2: ...ce have the following information available when contacting UAS UAS ORDER _________________________________________________________________ UNIT MODEL ________________________________________________________________ UNIT SERIAL _________________________________________________________________ SYSTEM ACCESSORIES ______________________________________________________________________________ ________...

Page 3: ...G 2 5 1 Normal Maintenance 2 5 2 Manual Cleaning of SMOG HOG Components 3 5 3 Manual Cleaning Methods 3 5 4 Appearance Of Components After Cleaning 4 5 5 Replacing Damaged Ionizer Wires 4 6 TROUBLESHOOTING 4 6 1 Tools Required 4 6 2 Perform Before Troubleshooting 4 6 3 Troubleshooting Procedures 5 6 4 Troubleshooting Table 8 7 REPLACEMENT PARTS 9 7 1 Replacement Parts List 9 8 FIGURES 10 12 9 WIRI...

Page 4: ...cessive accumulation which may result in flash over or fire damage The air cleaner should not be used for support of personnel or material Operate only in a safe and serviceable condition DANGER WARNING CAUTION C A U T I O N This is the safety alert symbol This symbol alerts you to potential hazards that can kill or hurt you and others All safety messages will follow the safety alert symbol and th...

Page 5: ...en using power pack 21 1216 and 5 6mA when using power pack 21 1234 In the event of a short circuit or other overload condition the power supply is designed to cause the high voltage to collapse When the voltage collapses the indicator light will blink When the overload is removed the voltage will automatically build up to normal and the indicator light will illuminate The power supply is self pro...

Page 6: ...ions Periodically momentary shorts will occur indicated by the flickering indicator light If the light is dim fails to glow or continues to flicker the components may need cleaning or servicing see Cleaning Instructions Section 5 2 and 5 3 5 MAINTENANCE AND MANUAL CLEANING 5 1 Normal Maintenance Once the unit is operational periodic maintenance is necessary to assure proper performance Follow a re...

Page 7: ...mation or some other undesirable condition Time is necessary for effective cleaning Soaking may seem slower and less effective than high impingement and or hot water above 180 F for cleaning action so personnel should be forewarned about using excessive pressure or temperature to shortcut the cleaning process High pressure or stream spray cleaning until plate edges are shiny is not effective Not o...

Page 8: ...kets are cleaned and intact 3 Component tracks are free of contaminant build up for unicell grounding 4 Module drain sump is cleaned and free flowing 5 Interior is free of extreme contaminant build up 6 Blower wheel is free of contaminant build up 5 5 REPLACING DAMAGED IONIZER WIRES 1 Remove the damaged wire from each spring Also replace spring if it is damaged 2 Loop one end of the new wire over ...

Page 9: ...high voltage The grounding procedure statement is not identified within the troubleshooting steps but should be performed The power pack total current output is limited to a maximum of 5 milliamps to assure personnel safety A high voltage probe is required to measure high voltages to the ionizer and collector circuits to the unit As an accessory a high voltage probe can be purchased for a multimet...

Page 10: ...ar itself by removing the unicell s from the module and then installing the unicell s back into the module Place MSH off connect all wires which have been disconnected and install unicell s b Indicator light is illuminated with the high voltage wire 7 connected to the collector cell connector on the power pack Place MSH off and connect high voltage wire 8 to the ionizer connector on the power pack...

Page 11: ... 11 8 KVDC If the ionizer voltage is below specifications refer step 2 Ionizer Conditions Causing Failure Low ionizer voltage will decrease the collector cell voltage but low cell voltage will not affect the ionizer voltage 4 Place the tip of the high voltage probe to the collector cell door feed through insulator bolt The high voltage measurement should be 5 0 to 7 3 KVDC If the collector cell vo...

Page 12: ...fective Unicell dirty Unicell malfunction Unicell installed backwards or upside down High voltage contacts dirty Loss of input voltage Unicell malfunctioning Power supply defective POSSIBLE SOLUTIONS Replace lamp Check connection to lamp Check power supply Replace air mover Clean foreign object from air mover Replace control switch Check connection at control switch Correct wiring Engage safety in...

Page 13: ...r 33Ͳ10116 33Ͳ10121 2 Unicell 02Ͳ10429Ͳ0001ͲS 02Ͳ1921ͲS 3 Mesh AfterͲFilter 33Ͳ10110 33Ͳ10120 4 Inlet Cone Ring 32Ͳ10053 32Ͳ10067 5 Motorized Impeller Fan 32Ͳ10052 32Ͳ10065 6 Ionizer Spring 36Ͳ0068 36Ͳ0068 7 Cell Spring 36Ͳ0012 36Ͳ0012 8 Cell Ground Spring 36Ͳ0016 36Ͳ0016 9 Interlock Switch 20Ͳ0005 20Ͳ0005 10 Filter Door Latch 39Ͳ10055 39Ͳ10055 11 Feed Through Insulator 37Ͳ0026 37Ͳ0026 12a Power P...

Page 14: ...10 Revised 03 15 SMOG HOG Electronic Air Cleaner Figure 1 Unit Components ...

Page 15: ...1 Revised 03 15 SMOG HOG Electronic Air Cleaner Figure 2 Drain Loop Assembly Figure 4 Replacing Ionizer Section Figure 3 Ionizer and Cell Voltage Measurement Figure 5 Grounding The Collector Cell Section ...

Page 16: ...12 Revised 03 15 SMOG HOG Electronic Air Cleaner Figure 6 Grounding the Ionizing Section Figure 7 Power Pack Input Output Connections ...

Page 17: ...13 Revised 03 15 SMOG HOG Electronic Air Cleaner MSH 05 ...

Page 18: ...14 Revised 03 15 SMOG HOG Electronic Air Cleaner MSH 05 M ...

Page 19: ...15 Revised 03 15 SMOG HOG Electronic Air Cleaner MSH 11 ...

Page 20: ...16 Revised 03 15 SMOG HOG Electronic Air Cleaner MSH 11 M ...

Page 21: ...17 Revised 03 15 SMOG HOG Electronic Air Cleaner MSH 11 w Optional ON OFF Switch ...

Page 22: ...18 Revised 03 15 SMOG HOG Electronic Air Cleaner MSH 05 1PH Interconnect MSH 11 1PH Interconnect ...

Page 23: ...19 Revised 03 15 SMOG HOG Electronic Air Cleaner MSH 05 3PH Interconnect MSH 11 3PH Interconnect ...

Page 24: ...dified except by a written instrument signed by the President or Vice President of Sales of UAS THIS WARRANTY IS EXCLUSIVE AND IN LIEU OF ALL OTHER WARRANTIES WHETHER WRITTEN ORALOR IMPLIED INCLUDINGANYIMPLIED WARRANTYOF MERCHANTABILITY FITNESS FOR A PARTICULAR PURPOSE OR NONINFRINGEMENT As Purchaser s exclusive remedy for any defects in the equipment UAS will exchange or repair any defective part...

Reviews: