PAGE 5 OF 6
1114 IH-4729
SEALING
PHYSICAL SEALS
For jurisdictions that use the physical sealing method,
the local weights and measures official or authorized
service agent must apply a security seal to prevent
tampering with the settings. Refer to the illustrations
below for sealing methods. (See Figures 2-3)
AUDIT TRAIL SEAL
For jurisdictions that use the audit trail sealing method,
the local weights and measures official or authorized
service agent must record the current configuration and
calibration event counter values at the time of sealing.
These values will be compared to values found during a
future inspection.
NOTE: A change to an event counter value is
equivalent to breaking a physical seal.
The audit trail uses two event counters to record
changes to configuration and calibration settings:
• The configuration event counter (CFG) will index
by 1 when exiting the menu if one or more of the
following settings are changed: Legal for Trade,
Calibration Unit, Capacity, Graduation, Power On
Unit, Zero Range, Auto Zero Tracking, Expanded
Mode, Count Mode, Kilogram Unit, Pound Unit, Gram
Unit, Ounce Unit, Pound Ounce Unit, Stable Only.
Note that the counter only indexes once, even if
several settings are changed. The configuration
event counter values range from CFG000 to CFG999.
When the value reaches CFG999, the count starts
over at CFG000.
• The calibration event counter (CAL) will index by 1
when exiting the menu if a Span Calibration, Linearity
Calibration or GEO setting change is made. Note
that the counter only indexes once, even if several
settings are changed. The calibration event counter
values range from CAL000 to CAL999. When the value
reaches CAL999, the count starts over at CAL000.
The event counters can be
viewed by pressing and
holding the
MENU
button.
While the button is held,
the display will show MENU
followed by AUd It.
1. Release the button when
AUd It is displayed to view
the audit trail information.
2. The audit trail information
is displayed in the format
CFGxxx and CALxxx.
3. Then the indicator returns
to normal operation.
OPERATION CONTINUED
Wire Seal
Paper Seal
Figure 2
Figure 3
CALIBRATION
MODE
CALIBRATION
MODE
CALIBRATION
MODE
CALIBRATION
MODE
CALIBRATION
MODE
CALIBRATION
MODE
CALIBRATION
MODE
CALIBRATION
MODE
CALIBRATION
MODE
MAINTENANCE
CAUTION! Disconnect the unit from
the power supply before cleaning.
CLEANING
• The housing may be cleaned with a cloth
dampened with a mild detergent if necessary.
• Do not use solvents, chemicals, alcohol, ammonia
or abrasives to clean the housing or control panel.