PAGE 2 OF 3
0515 IH-3478
OPERATION
SET UP MENU
TURNING THE SCALE ON/OFF
• To turn the scale on, press the
ON/OFF
button.
• To turn the scale off, press the
ON/OFF
button.
CHANGING UNIT OF MEASURE
• Press the
MODE
button repeatedly until the desired
unit of measure is displayed and release.
CALIBRATION
1. Press and hold
ZERO
while turning on the scale. Do
not release
ZERO
until the OFFSET value is displayed.
2. Press
ZERO
again. After "Zero" is displayed, the
required calibration weight of
400g
is displayed.
3. Place the corresponding calibration weight on the pan.
4. Press
ZERO
to calibrate.
CAL
will be displayed on
the screen.
5. The calibration procedure is completed when
PASS
is displayed on the screen. Remove the weights.
6. Press
TARE
to return to normal weighing mode.
WEIGHING
1. Place item to be weighed on the scale platform.
2. Weight will be shown in the display.
COUNTING
1. Press
MODE
to select “PCS.”
2. Press
N/G
to choose the sampling amount:
5=10
10 pieces
5=20
20 pieces
5=50
50 pieces
5=100
100 pieces
3. Place a sample on the weighing pan and press
MODE
.
CAL
is displayed momentarily before the
display reverts to the sample size.
4. Remove the samples. Place additional items on the
platform. The scale will begin to count.
Note: The sampling result is retained when the
unit is turned on.
TARE
1. Place container on the scale.
2. Press the
TARE
button. Scale will zero.
3. Add items to be weighed or counted into the
container.
SET UP MENU NAVIGATION
1. To enter the menu, press and hold
ZERO
while
turning on the scale. Do not release
ZERO
until the
OFFSET value is displayed.
2. Press
MODE
, the screen displays
Ad
.
3. Press
N/G
to step through all the settings.
4. Press
MODE
to initial the displayed setting, and then
press
N/G
to step through internal parameters.
5. Press
MODE
to save the setting. Press
TARE
to exit
without saving.
6. Press
ZERO
to return to weighing mode.
BACKLIGHT
LiGHT
: Activation of the backlight.
Auto Auto on with items greater than
0.9g placed on the weighing pan
ON
Backlight on
OFF
Backlight off
AUTO POWER OFF
AUoFF
: Set the period of inactivity before the scale
automatically powers off.
NO
Do not power-off
5
5 minutes
10
10 minutes
15
15 minutes