background image

PAGE 7 OF 8 

0313 IH-3827

DYNAMIC/ANIMAL WEIGHING 

CONTINUED

AUTOMATIC MODE

The balance returns to the Ready condition when the 

animal is removed from the pan. If the Function button 

is pressed before the animal is removed from the pan, 

the primary display shows the actual weight and the 

secondary display alternates between ANIMAL and 

CLEAR PAN until the animal is removed from the pan.  

The balance then returns to the Ready condition.

SEMIAUTOMATIC AND MANUAL MODES

Remove the animal from the pan and press the 

Function button to return to the Ready condition. If the 

Function button is pressed while the animal is on the 

pan, the primary display shows the actual weight and 

the secondary display alternates between ANIMAL and 

CLEAR PAN. The balance returns to the Ready condition 

when the animal is removed from the pan. The Animal 

Weigh level and mode settings are made in the Mode 

menu.

ADDITIONAL FEATURES

WEIGH BELOW

The Adventurer Pro balance is equipped with a weigh 

below hook.

NOTE: Before turning the balance over,  
install the cone cover to prevent damage  
to the load cell.

To use this feature, remove power from the balance 

and remove the protective cover for the weigh below 

opening.

The balance can be supported using lab jacks or any 

other convenient method. Make sure the balance is 

level and secure. Use a string or wire to attach items to 

be weighed. 

 

 

 

 

BATTERY OPERATION

Models with the battery operation feature will display 

the battery change indicator as shown below.  

If batteries are installed and AC power is not present, 

the balance operation is affected as follows:

•  The LCD backlight is normally off to conserve power. 

The backlight turns on when the display becomes 

unstable or when a button is pressed. The backlight 

turns off 5 seconds after the display becomes stable or 

after 5 seconds with no button presses. The backlight is 

always off if the Backlight menu setting is OFF.

•  Auto Off is set to 5 minutes. If continuous operation is 

desired, change the Auto Off menu setting to Off.

•  The battery charge level is annunciated by means 

of a 3 segment battery symbol on the display as 

follows: 

CLEANING

If cleaning is required, use a soft cloth dampened with 

water and a mild detergent. Do not allow liquids to enter 

the balance. Do not use harsh chemicals to clean the 

balance as the finish may be damaged.

Full charge

2/3 charge
1/3 charge
Replace batteries

(Flashing)

OPERATION CONTINUED

Summary of Contents for ADVENTURER PRO Series

Page 1: ...igure 1 3 For applicable Adventurer Pro models place the Wind Ring over the pins located on the perimeter of the balance INSTALLING COMPONENTS Use the illustrations and instructions below to identify and assemble your Adventurer Pro model with its com ponents All components must be installed before using the balance ASSEMBLY OF MODEL H 3827 1 Insert the Sub platform on the Mounting Cone located in...

Page 2: ... front of the balance and the small notch on the bottom edge of the door is lined up with the tab that protrudes from the side of the balance housing See Figure 3 Insert the top edge of the glass door into the recessed area under the top cross member of the Draft Shield framework Slide the notch on the bottom edge over the tab and slip the door in place 5 The door may now be opened by sliding it t...

Page 3: ...ING BUTTON CONTROL FUNCTIONS SETTING THE BALANCE TO ZERO Remove the load from the pan and press the Zero button to set the display to zero When the weighing pan or platform is empty the Center of Zero indicator turns on when the measurement is within 1 4 d of the zero setting TARING Taring refers to the action of allowing for the weight of a container so that only the weight of objects held in the...

Page 4: ...nu that appears on the secondary display blinking The sub menu name now appears on the primary display and the first menu item in the sub menu now appears on the secondary display blinking SELECT THE MENU ITEM Press the No button to select the next setting or the Back button to select the previous setting in the loop Press the Yes button to select and save the setting on the secondary display blin...

Page 5: ... press the Function button This example illustrates 50 grams as the reference weight 2 Remove the reference weight and place the item to be compared on the pan The balance indicates the actual percentage on the primary display The secondary display indicates the weight in the selected measuring unit 3 Pressing the Function button momentarily displays the reference weight on the secondary display C...

Page 6: ... entered The secondary display shows the previous setting or all zeros if no limit was stored blinking Press the Yes button to accept this setting Press the No button to edit this setting The Under limit is the first to be edited followed by the Over limit EDITING LIMITS The limits are edited one digit at a time The digit being edited is blinking Its value can be incremented by pressing the No but...

Page 7: ...ure remove power from the balance and remove the protective cover for the weigh below opening The balance can be supported using lab jacks or any other convenient method Make sure the balance is level and secure Use a string or wire to attach items to be weighed BATTERY OPERATION Models with the battery operation feature will display the battery change indicator as shown below If batteries are ins...

Page 8: ... reference weight Eliminate vibration and drafts Error 8 1 Weight reading exceeds Power On Zero limit Clear platform Error 8 2 Weight reading below Power On Zero limit Put platform on a balance Error 8 3 Weight reading exceeds overload limit Clear platform Error 8 4 Weight reading below overload limit Put platform on balance Error 9 0 Internal fault Return balance for service Error 9 5 Production ...