The preset reminder alarm is ringing
Press any button to turn off the alarm or wait for the alarm to
be turned off automatically in 30 seconds.
Temperature is TOO HIGH during the test procedure.
Large variation may occur between results due to high or
low temperature. Move to an environment
(50°F ~ 104°F or 10°C ~ 40°C) and wait for 30 minutes before
re-testing.
Battery power is LOW. Meter will only provide
proximately 10 more measurements.
Replace with 2 AAA batteries.
1. An error message indicates that the hematocrit of
the blood sample is too low for HT100-B system to
perform a test.
2.Use control solution to check the meter and test
strips.
1.Follow the instructions and try again with a new test strip.
Please contact customer service if the problem persists.
2.Follow the instructions to enter control solution mode, and
try again with a new test strip.
Temperature is TOO LOW during the test procedure.
Large variation may occur between results due to high or
low temperature. Move to an environment (50°F ~ 104°F or
10°C ~ 40°C) and wait for 30 minutes before re-testing.
Display
Description
Action
An error message indicates that the hematocrit of
the hematocrit of the blood sample is too high for
HT100-B system to perform a test.
Follow the instructions and try again with a new test strip.
Please contact customer service if the problem persists.
29