11
STARTING AND OPERATING
Set values
Actual values
Entry keys
Left (-)
Right (+)
Starting the Machine
Switch on the machine with the
main switch
(located at the rear of the unit).
All Indicators on the front panel will read “0”.
Time
The sieving time is shown in minutes (99 minutes maximum) and appears in the indication field of
set values/actual values. The sieving time will countdown to “0”. Without setting the sieving time
(set point 00) the machine operates continuously.
The sieving time can be fed by one pair of entry keys for each function. The left key decreases,
while the right key increases the test time. When pressing one of the keys for a longer moment, the
numbers pass quickly.
Analysis Settings
There are two amplitude analysis settings
.
One for
coarse
and the second for
fine material
.
The Model Ro-Tap
®
E features an intermittent operation. Every 10 seconds,
the machine will pause for 3 seconds, allowing the sample to settle.
Analysis Guide: If the smallest particle is 1 mm (USA #18) or greater in size,
use the COARSE analysis control setting. For all other applications,
use the FINE setting.
Each material application is different. Local analysis may be
required to determine which
control setting will produce the best results.
Start/Stop
To start the machine, press the START button. As the unit runs, the remaining analysis time
will be shown in the display window. To stop the machine, press the START button again.