10. The measurement Data Storage
At the end of each test, the test data can be automatically saved by machine,
and only the last 20 sets of test data are retained for easy access to view, print,
etc.
Button (command) Step description Display
Standby
0.000 g
Long press [PRINT] Press 3 seconds
NOTES
The flash save number and data is the last time of moisture measurement
Short press [↓] To display the next group of measure data
Short press [↑] To display the last group of measure data
Long press [PRINT] To print the current data
Long press [
ON/OFF
] Return to weighing status
11. Maintenance and troubleshooting
11.1 Meaning of signal
O Err
This signal remind user that machine was not covered
T Err
Remind user that temperature not increase after start to heat
Lo
Remind that the test sample weight is less than the minimum of
required sample weight
CAL 0
Remind that machine will calibrate on zero ( no need to put any
weight mass on pan)
C 100
Remind user to put 100g weight mass on pan for calibration
C EnD
The calibration was successful
C Err
The calibration failed
If the fault message cannot be resolved or does not describe the problem with your moisture
analyzer, please call our customer service department or your local authorized service agent
or company.
11.2 Maintenance of machine
Cleaning: For accurate measurement results, it is recommended to clean the temperature
sensor and heating silo regularly.
Before cleaning, machine must be unplugged, clean carefully, so as not to damage the
instrument, please use a hairless (velvet) soft cloth and neutral detergent cleaning, and never
use corrosive detergent. Do not allow the detergent to flow into the inside of the instrument.