Calibration Procedure
21
Calibration is a function to optimize the monitor display color output. Calibration
program can be used for all TVLogic Monitors. For a detailed calibration procedure,
please refer to the following.
Before starting
Fig 1: Connecting to the serial Port
1. Connect the Monitor’s RS232 port to the PC’s serial port using the 9 pin
serial cable. Only TX, RX and GND must be connected. Therefore, the
additional cable provided with the initial serial cable must be connected (TX,
RX, GND connector cable).
RS-232
USB & Serial
2. Connect the measuring device and the PC with a serial cable or a USB. Use an
appropriate cable depending on the type of measuring device.
3. When you connect to X-Rite, CA-210, CS-200 or Eye-One, please use a USB to
connect the measuring device and the PC. Device driver provided by the
measuring device needs to be installed when connecting via USB.
4. Locate the measuring device at the display’s center.
5. Set SDI-A as the monitor’s input source.
6. You can get more accurate results if you pre-heat the monitor for 30 minutes.
Summary of Contents for XVM-175W
Page 1: ...17 Multi Format LCD Monitor XVM 175W Multi Format LCD MONITOR Service Manual 1 ...
Page 3: ...Part Name Function 3 XVM 175W Front XVM 175W Rear ...
Page 4: ...Drawing ...
Page 5: ...Exploded View ...
Page 8: ...PCB Block Diagram 8 ...
Page 37: ...Circuit Diagram SB IC1 IC5 IC2 IC3 U3 IC6 ...
Page 38: ...Circuit Diagram EB U5 U6 U1 U3 U2 U7 U8 IC12 CON6 CON2 ...
Page 39: ...Oscilloscope BCJ_FPC02 SD_Y_A5 DATA_IN_B7 ...
Page 40: ...Oscilloscope TXOUT DATA_OUT32 FRN0 ...
Page 41: ...Oscilloscope EM_AUDIO_L I2S_SDI LEFT_0 ...
Page 42: ...Oscilloscope V556_CLK CLK_A CLK_IN1 ...
Page 43: ...Oscilloscope PHS PCLK_OUT DCLK1 ...
Page 50: ...Packaging ...