23
5
Installation Instructions
Lifting and Carrying
Cautions!
Before moving the autoclave, make sure that the electric cord is
disconnected from the power, and there is no pressure in the chamber.
Drain the water from the reservoir (see draining the reservoir, 8.9)
Do not drop the device!
To avoid injuries, lifting and carrying should be done with at least two people or
by using a fork-lift or any other mechanical aid.
Device Placement and Operating Conditions
1. The autoclave is intended for indoor use only.
2. Check and verify that the counter carrying the autoclave is a rigid and
leveled surface and can carry a load of 165 lbs (75kg).
Caution
!
The device is not designed for
use on any standard slide out shelf. If it is
necessary to use a slide out shelf, it must be
tested and/or rated for 165 lbs (75Kg) or more.
3. Check and verify that the dimensions of the surface of the counter are at
least 21"(55cm) x 23"(60cm).
4. Keep the back and the sides of the autoclave approximately 10cm away
from the wall to allow ventilation and facilitate the device disconnection.
5. If placed in a cabinet, verify that the rear of the cabinet is open to allow
ventilation.
Caution! Insufficient space for
ventilation may result in malfunction or
damage due to overheating.
6. It is recommended that enough space be left around the autoclave to
give a technician access for servicing the machine.
7. Check and verify that the ambient temperature range is 41ºF (5ºC)-
104ºF(40ºC), it is preferable not to exceed 86 ºF (30ºC).