Page 51
Printer output
Description
Operator
:
____________
To be filled in manually by operator
Time:
12:14:47
Time sterilization cycle ended
Status: Cycle
Ended
00:24:43
101.3
099.7
Cycle finished time
D 00:31:23
101.9
023.4
The time, temperature and pressure during drying
D 00:28:23
106.2
109.7
The time, temperature and pressure during exhaust
E 00:23:43
107.0
107.4
The time, temperature and pressure during
exhaust
E 00:22:08
134.5
311.9
The time, temperature and pressure during
exhaust
CLK 2:
12:12:10:00
Sterilization Process End time as registered by two
clocks
CLK 1:
12:12:10:00
S 00:22:07
134.5
311.6
The time, temperature and pressure during
sterilization
S 00:22:06
134.5
311.6
The time, temperature and pressure during
sterilization
S 00:21:06
134.6
311.0
The time, temperature and pressure during
sterilization
S 00:20:06
134.5
310.1
The time, temperature and pressure during
sterilization
S 00:19:06
134.8
311.1
The time, temperature and pressure during
sterilization
S 00:18:06
134.5
315.8
The time, temperature and pressure during
sterilization
CLK 2:
12:08:08:00
Sterilization Process Start time as registered by two
clocks
CLK 1:
12:08:08:00
H 00:18:04
134.4
315.1
The time, temperature and pressure during
heating
H 00:16:35
128.9
268.4
The time, temperature and pressure during
heating
H 00:13:35
116.3
110.9
The time, temperature and pressure during
heating
A 00:08:48
089.6
077.1
The time, temperature and pressure during air
removal
A 00:05:48
091.3
049.4
The time, temperature and pressure during air