TURTLE
TOUGH
®
26
ISE
Measurement
ISE
Setup Parameter
Configuration
6
N/A
Sensor Type
5
P18
Default Node
5
P02
Default Baudrate
19,200
P03
Default Output Type
4-20mA
P04
Default Polarity
non-inverted
P05
Default Low Whole
44
P06
Default Low Decimal
44
P07
Default Hi Whole
22
P08
Default Hi Decimal
22
P09
Days to Recalibrate
14
P17
CHANGE VALUE BELOW TO MATCH P18 WHEN SMART RTU ISE SENSOR IS CONNECTED
if P16 Value is: 19.00
Integer Limits
Engineered pION Limits
0
-2.000
18,000
16.000
% of Full Range
Engineered pION Units
RTU Integer
ppm units
0.00%
-2.000
0
19000000
5.56%
-1.000
1000
1900000
11.11%
0.000
2000
190000
16.67%
1.000
3000
19000
22.22%
2.000
4000
1900
27.78%
3.000
5000
190
33.33%
4.000
6000
19
38.89%
5.000
7000
1.9
44.44%
6.000
8000
0.19
50.00%
7.000
9000
0.019
55.56%
8.000
10000
0.0019
61.11%
9.000
11000
0.00019
99.75%
7.978
ppm Low Set
0.00200
P06/P07
66.42%
4.978
ppm High Set
2.00000
P08/P09
% FULL RANGE COMPUTED FOR PPM
VALUES ENTERED TO THE RIGHT
CHANGE ppm VALUES ABOVE TO DESIRED VALUES
FOR LOW & HIGH SETPOINTS
44.44%
6.000
Default Low Setpoint in pION
22.22%
2.000
Default High Setpoint in pION
Note 1:
Low and high analog setpoints should be at least 1,000 MODBUS RTU steps apart.
Note 2:
0 ppm not a valid number for low setpoint since there exists no corresponding pION value.