Pronova-O2 preparation

Preparation of Pronova-O2™ for Patient Placement

This section describes preparations and considerations to prepare the 
Pronova-O2™ surface for placement. 

1.    Plug into properly grounded outlet.
2.    Gather InteliDerm™ Face Pack, Chest Wedge, and Abdominal
       Sling, place on surface and zero scale.
3.    Clear History: Select Begin New Patient from settings menu.
4.    Ensure Critical Line Management System is secured and installed on
       Head Hoop.
5.    Open Head Hoop.
6.    Lower and stow Head Side Supports, pull knob to rotate and lower.
7.    Remove Pelvic Pads from Pelvic Prone Support.
8.    Lower and stow Prone Surface Supports (see images).

1.    Unbuckle and rotate surface to attach to prone support arm.
2.    Pull knob on left side bottom of prone arm to lower parallel
      to surface.
3.    Press Prone Support to stow beneath surface.

9.    Unbuckle Shin Slings.
10.  Lower and stow Side Support Panels (see images).

1.    Press button on side of frame to pull outward to edge
      of surface.
2.    Pull knob under center of panel and lift handle to rotate
      outward and down beneath surface.

11.  Lower and stow Foot Rests.
12.  Lower and stow Inner Leg Support.
13.  Lift Accessory Rack.
14.  Lock brakes.
15.  Ensure Lock Pin is in locked position.

Summary of Contents for Pronova-O2

Page 1: ...QUICK REFERENCE GUIDE For Clinical or Technical Support Call 1 855 ASK TURN 1 855 275 8876...

Page 2: ...vary Contact Turn Medical to request product education training or a copy of the IFU at 1 855 ASK TURN 1 855 275 8876 Indications for Use The Pronova O2 Automated Prone Therapy System is designed to...

Page 3: ...ive ETT securement methods Secure patient s hair to prevent catching during rotation on surface Remove any piercings to reduce risk for skin breakdown Properly dress and secure any skin injury or surg...

Page 4: ...pports pull knob to rotate and lower 7 Remove Pelvic Pads from Pelvic Prone Support 8 Lower and stow Prone Surface Supports see images 1 Unbuckle and rotate surface to attach to prone support arm 2 Pu...

Page 5: ...4 Pronova O2 preparation 8 Lower and stow Prone Surface Supports 10 Lower and stow Side Support Panels...

Page 6: ...g securement straps 9 Secure Foot Rests 10 Raise secure and adjust Side Supports see images 1 Use handle on center of panel to raise to surface 2 Press button on side of frame and push inward to secur...

Page 7: ...roperly secure patient on surface Patient may shift during rotation Re assess securement often and tighten side supports against body as needed Adjustment can be made at any point in rotation 16 Raise...

Page 8: ...Initiate Therapy...

Page 9: ...on Checklist 7 Press and hold Prone Patient 8 Once 0 prone is reached consider placing patient in Reverse Trendelenberg using Hand Control or Home Screen 9 Initiate Prone Rotational Therapy Refer to P...

Page 10: ...est At Angle Return to Supine Position The following section reviews the steps to Return to the Supine Position 1 Press Stop if surface is in rotation or ensure Lock Pin is in unlocked position 2 Lowe...

Page 11: ...if surface is in rotation or ensure Lock Pin is in unlocked position 2 Press Prone Therapy or Supine Therapy tab 3 Input Verify Settings 4 Press Continue 5 Verify Checklist 6 Press Start Therapy Obser...

Page 12: ...surface to 0 Supine 4 Monitor all invasive lines for rotation 5 Engage Lock Pin in Locked position 6 Place CPR board under patient for proper compressions 7 Initiate CPR procedures per hospital facil...

Page 13: ...3 or more clinicians 2 to support frame 1 to release manual CPR lever 2 Place CPR board and close panels if in Prone Position 3 Determine direction of rotation to Supine Position toward ventilator opp...

Page 14: ...13 notes Notes...

Page 15: ...14 notes Notes...

Page 16: ...Copyright 2021 Turn Medical All rights are reserved A copy of patents and or patents pending can be found at Turnmedical com NOTICE 2021 Turn Medical All rights are reserved To support the intended us...

Reviews: