TD-700 Laboratory Fluorometer Operating Manual
Page 19
3.
Direct Concentration Calibration:
Direct Concentration Calibration,
only available in the Multi-Optional Mode, is a multi-point calibration in
which up to five standards and a blank are read. The software uses
these points to set the optimal instrument range and sensitivity, and to
calculate the direct concentration of unknowns. In a multi-point
calibration, the instrument generates a calibration curve for superior
accuracy. The TD-700 will display the actual concentration of your
samples. The display units are user-selectable and chosen in the
setup menu.
Note: Fluorescence measurements are affected by factors such as
temperature, linearity, and instrument drift. Before calibrating or
running samples for the first time, we recommend reviewing Appendix
1.