Maintenance/Battery Replacement 

Maintenance: 

 

To improve performance and avoid transferring soil 
borne diseases, clean the sensor tip by rinsing in 
alcohol for 5 - 10 minutes. 

 

Replace all batteries if low battery indicator 
appears, or if readings are faint or unstable. 

 

Store the probe sensor dry. 

Important:  Do not touch the sensor tip with your 
fingers.  The oils on the skin will affect to probe’s 
measurement accuracy. 

 

Changing Batteries: 

1.

 

Open battery compartment lid (located near the LCD screen). 

2.

 

Remove and replace the old batteries.  Note polarity shown in battery compartment.  
When inserting new batteries, take care to place them over the white removal tab.  This 
will make future battery removal much easier. 

3.

 

Recalibrate meter after every battery change. 

Specifications 

 

Readout 

LCD digital display 

EC Range 

0.00 - 19.99 mS/cm 

Temperature Range 

-5 - 55 C 

Accuracy 

EC: +/-2% Full Scale 

Temperature 

+/0.5 C 

EC Resolution 

0.01 mS/cm 

Operating Environment 

0 to 50C 

Calibration 

1-point with 2.76 mS/cm standard 

Temperature Compensation 

Automatic form 0 to 50C 

Power 

Four LR44 1.5V alkaline batteries 

Appendix 1 

Classification of Irrigation Water

 

 

Excellent 

EC < 0.25 mS/cm 

Good 

0.25 mS/cm < EC < 0,75 mS/cm 

Permissible 

0.75 mS/cm < EC < 2.0 mS/cm 

Doubtful 

2.0 mS/cm < EC < 3.0 mS/cm 

Unsuitable 

EC > 3.0  mS/cm 

Categories of irrigation water quality based on electrical conductivity.  (USDA Tech Bulletin 
962)

 

 
 
 