13
Maintenance and Service
Note: if the failure of oxygen generator cannot be eliminated,
please contact the supplier or manufacturer
The oxygen
generator works, but
there is little or no air
output at the oxygen
inlet
1
)
Kinked or blocked
oxygen pipeline
1)Straighten out the
pipeline
2
)
The rubber plug of
the humidification
bottle is not covered
tightly or is missing
2
)
Retighten
3)Blocked
humidification water
tank
3
)
Clean the
humidification water
tank
4
)
Filter dirty
4
)
Replace filter
5
)
Contact supplier or manufacturer if persistent
The oxygen
generator has power
supply, but the
machine does not
work, and the display
shows "fault"
1)Compressor
overheating
protection, shutdown
1
)
Shut down and
restart the machine
after it cools down
2
)
If this condition persists, stop using the
machine and contact the equipment supplier or
manufacturer immediately