background image

1

 Introduction

  

The TIA-952 Optical to Electrical Converter is a convenient 

fi

 ber optic detector/ampli

fi

 er combination 

that mounts directly on the input of an oscilloscope, digitizer, or other readout device.  With a band-
width of 800 MHz, it accurately provides an electrical replica of the optical signal presented to it. It is 
fully capable of driving a 50 ohm cable terminated in its characteristic load. 

The InGaAs version of the TIA-952 has a transimpedance of 500 

Ω

  plus a post ampli

fi

 er with select-

able gains of  1 or 5. Thus the overall responsivity ranges from approximately 450 V/W to 2250 V/W 
at the peak of the detector response curve and when terminated in a 50 ohm load.

The Silicon version of the TIA-952 has a transimpedance of 250 

Ω

  plus a post ampli

fi

 er with select-

able gains of  1 or 5. Thus the overall responsivity ranges from approximately 225 V/W to 1125V/W 
at the peak of the detector response curve and when terminated in a 50 ohm load.

The unit is powered by a universal wall mount power supply.  The supply accepts mains voltages 
from 95 to 260 V.A.C., 47-60 Hz. Four mains plugs are provided that are compatible with standards 
used in North  America, Great Britain, Continental Europe and Australia.

 Unpacking and Inspection 

Prior to shipment this instrument was inspected and found to be free of mechanical and electrical 
defects.  Upon acceptance by the carrier he assumes responsibility for its safe arrival.  After unpack-
ing, examine the unit for any evidence of shipping damage.  Should you receive this instrument in a 
damaged condition, apparent or concealed, it must be noted on the freight bill or express receipt and 
signed by the carrier’s agent. Failure to do so could result in the carrier refusing to honor the claim. 
Upon 

fi

 ling a claim TTI should be noti

fi

 ed.    

Calibration - This is a qualitative measurement device. No calibration is required or necessary

.

Summary of Contents for TIA-952

Page 1: ...TIA 952 Optical Electrical Converter Terahertz Technologies Inc 169 Clear Road Oriskany NY 13424 315 736 3642 FAX 315 736 4078 E mail sales terahertztechnologies com 2 2012 Operating Instructions...

Page 2: ...Introduction 1 Unpacking and Inspection 1 Powering the Unit 2 Controls 2 Operating Considerations 3 Spectral Response 3 Specifications 4 Limited Warranty 5 Repair Information 5 Contents...

Page 3: ...om approximately 225 V W to 1125V W at the peak of the detector response curve and when terminated in a 50 ohm load The unit is powered by a universal wall mount power supply The supply accepts mains...

Page 4: ...in North America Europe Great Britain or Australia Simply snap on the appropriate mains connector and plug the 2 1 mm coaxial plug into the receptacle located just under the optical input connector T...

Page 5: ...pedance and further multiplied by the second stage gain For example the sensitivity of the unit at a wavelength of 1550 nm would be 0 9 A W x 500 V A x 5 2 250 V W When using the TIA 952 mounted on an...

Page 6: ...5 125um multi mode Output Offset Voltage N A Volts Noise Level 9 5 pW root Hz at peak responsivity Maximum Output Voltage Low gain 2 0 V pk pk High Gain 4 V pk pk Into 50 Ohms Power Requirements Supp...

Page 7: ...THE ONLY WARRANTY EX PRESS OR IMPLIED GIVEN BY TTI I E THERE IS NO WARRANTY OF FITNESS FOR A PARTICULAR PURPOSE TTI HEREBY DISCLAIMS ANY EXPRESS OR IMPLIED WARRANTY OTHER THAN THE WARRANTY IN THE FIRS...

Reviews: