![TSI Instruments N95-Companion 8095 Operation And Service Manual Download Page 20](http://html1.mh-extra.com/html/tsi-instruments/n95-companion-8095/n95-companion-8095_operation-and-service-manual_1178269020.webp)
Chapter 2
12
Installing the Probe
1.
Select the mask you wish to test.
2.
Slide the sampling probe onto the piercing tool.
As the pointed end of
the piercing tool is very sharp, be extremely careful when handling it!
3.
Choose a location on the mask that is in front of the person's
nose/mouth region. Avoid seams and folds in the mask.
4.
Pierce the mask at the selected location, using the piercing tool with the
loaded sampling probe.
Be sure to pierce the mask from the inside!
Sampling
Probe
Push Nut
Positioned on
Anvil Tool
Sampling
Probe Pushed
Through Mask