6-4
Hydronic Manometer Model HM685
T e m p e r a t u r e
Temperature readings can display in °C or °F.
At the Select Units menu, highlight Temperature and press the
Enter
key to select the desired setting.
T i m e C o n s t a n t
The current value of the time constant is adjustable. Available settings
include 1, 5, 10, 20, and 30 seconds.
Notes:
The time constant is the period over which the manometer averages
measurements. Example: with the time constant equal to ten (10)
seconds, the displayed reading represents the average of
measurements taken over the previous ten (10) seconds.
Increasing the time constant will serve to improve measurement
stability, particularly when measuring systems with fluctuating
pressures.
Summary of Contents for HM685
Page 1: ...HYDRONIC MANOMETER MODEL HM685 OWNER S MANUAL P N 6006820 REVISION D FEBRUARY 2016...
Page 2: ......
Page 10: ...viii Hydronic Manometer Model HM685 This page intentionally left blank...
Page 12: ...x Hydronic Manometer Model HM685 This page intentionally left blank...
Page 16: ...xiv Hydronic Manometer Model HM685 This page intentionally left blank...
Page 20: ...1 4 Hydronic Manometer Model HM685 This page intentionally left blank...
Page 34: ...2 14 Hydronic Manometer Model HM685 This page intentionally left blank...
Page 40: ...3 6 Hydronic Manometer Model HM685 Entered Temperature...
Page 60: ...5 14 Hydronic Manometer Model HM685 This page intentionally left blank...
Page 72: ...7 4 Hydronic Manometer Model HM685 This page intentionally left blank...
Page 78: ...8 6 Hydronic Manometer Model HM685 This page intentionally left blank...
Page 92: ...B 2 Hydronic Manometer Model HM685 This page intentionally left blank...