Installing the Heated Inlet 
(854041) 

1.  Attach heated inlet control module to the front of the 

Photometer. This can be done using the 2x 6-32 x 
5/16 Socket Head Cap Screws and 2x flat washers. 
Use a 7/64” ball driver.  

 

2.  Push out elastic plug and install RH/Temp Sensor 

sealing gland. Removable nut should be installed on 
the inside of the enclosure. 

 

 

W A R N I N G

 

Water gasket on gland must form a good 
seal with the exterior of the enclosure to 
ensure a waterproof seal. 

3.  Insert RH/Temp 

Sensor through 
gland and tighten 
gland nut. Extend 
probe ~2.75” (7 cm) 
beyond nut. 

 

4.  Attach the sun 

shield to the top of 
the RH/Temp 
Sensor as shown. 
Sun Shield should 
not touch sintered 
cap of RH/Temp 
Probe. 

 

5.  Attach RH/Temp 

Sensor cable to 
heated inlet control 
module (J3). 

 

6.  Attach the two 

larger (gray) 
connectors to side 
of the photometer. 

 

7.  Attach the two 

smaller (white) 
connectors to the 
Power

 

/

 

Communication 
inputs (J1 and J2) 
of the heated inlet 
control module. 

 

8.  Install the water trap bottle onto the side of the inlet 

as shown. 

NOTE 

For best results, use PTFE Thread Sealant 
Tape on the water bottle threads. 

 

Summary of Contents for DustTrak 854041

Page 1: ...___ TSI and TSI logo are registered trademarks of TSI Incorporated DustTrak is a trademark of TSI Incorporated HEATED INLET SAMPLE CONDITIONER ACCESSORY 854041 FOR DUSTTRAK ENVIRONMENTAL MONITORS INSTALLATION GUIDE ...

Page 2: ......

Page 3: ...owing table are present 2 Contact TSI immediately if items are missing or broken 3 Additional items may be included if you ordered accessories or spare parts Qty Item Description Reference Picture 1 Heated Inlet Column 1 RH Temp Sensor 1 RH Temp Sensor Gland 1 Heated Inlet Control Module 1 Omni Direction Inlet Qty Item Description Reference Picture 1 Water Trap 1 Power and Communication Cable 1 Co...

Page 4: ... ensure a waterproof seal 3 Insert RH Temp Sensor through gland and tighten gland nut Extend probe 2 75 7 cm beyond nut 4 Attach the sun shield to the top of the RH Temp Sensor as shown Sun Shield should not touch sintered cap of RH Temp Probe 5 Attach RH Temp Sensor cable to heated inlet control module J3 6 Attach the two larger gray connectors to side of the photometer 7 Attach the two smaller w...

Page 5: ...umn 12 Attach the inlet to the column 13 Install O ring and route cable through the mounting ring Pull excess cable through to prevent pinching of cable when inlet column is installed 14 Set inlet column on top of photometer mounting ring rotate slowly until alignment pin mates with the hole in the bottom of heated inlet column 15 Slide column retainer ring down heated inlet column 16 Hand tighten...

Page 6: ...nscrew the top knurled ring collar from the lower mounting collar 2 Inside the Enclosure disconnect the Heated Inlet cable note that the connector has a lock release on it 3 Remove the Ferrite Clamp Assembly Note that the clamp is released by inserting and twisting a small flat blade screwdriver Also for purposes of re assembly note that the cable twists two times through the Ferrite 4 Carefully p...

Page 7: ...escribes the status of the heated inlet Solid Green Inlet temp is 1 C below set point and controlling to maintain set point Blinking Green Inlet temp is between 1 to 5 C below set point and controlling to improve Blinking Red Inlet temp is more than 5 C below set point and controlling to improve This will occur when unit is first turned on and coming to temperature Solid Red Sensor unplugged or ha...

Page 8: ... more information USA Tel 1 800 680 1220 UK Tel 44 149 4 459200 France Tel 33 1 41 19 21 99 Germany Tel 49 241 523030 India Tel 91 80 67877200 China Tel 86 10 8219 7688 Singapore Tel 65 6595 6388 P N 6008577 Rev D 2018 TSI Incorporated Printed in U S A ...

Reviews: