The Model 5825 requires very little maintenance to keep it performing
well.
Recalibration
To maintain a high degree of accuracy in your measurements, we
recommend that you return your Model 5825 to TSI for annual
recalibration. Please contact one of TSI’s offices or your local
distributor to make service arrangements and to receive a Return
Material Authorization (RMA) number. To fill out an online RMA form,
U.S. & International
TSI Incorporated
The Model 5825 can also be recalibrated in the field using the
CALIBRATION menu. These field adjustments are intended to make
minor changes in calibration to match a user's calibration standards.
The field adjustment is NOT intended as a complete calibration
capability. For complete, multiple-point calibration and certification,
the instrument must be returned to the factory.
Cases
If the instrument case or storage case needs cleaning, wipe it off with
a soft cloth and isopropyl alcohol or a mild detergent. Never immerse
the Model 5825. If the enclosure of the Model 5825 or the AC adapter
becomes broken, it must be replaced immediately to prevent access
to hazardous voltage.
Storage
Remove the batteries when storing the unit for more than one month
to prevent damage due to battery leakage.