Chapter 4
56
To access the filters open the top cover of the pump module. The two HEPA
filters are identified in the figure below. The two filters can be replaced by
disconnecting the soft tubing between the filters, pump, and the casing
connectors.
C a u t i o n
When replacing the HEPA filters, make sure they are
oriented in the correction direction as shown in the picture
above.
Storage Precautions
When storing the DustTrak monitor for more than 30 days, you should charge
and remove the batteries. This prevents damage due to battery leakage.
This instrument must be stored in a location where the temperature remains
between
–20 and 60°C (–4 and 140°F).
User Replaceable HEPA Filters