Calibration
Calibration adjustment is permanent and should only
be performed using accurate calibration weights that
match the requirements for your scale (see the chart
on back). Do not use coins or other household items
to calibrate.
1.
Prepare the required weight for your scale. (see
chart on back)
2.
Find a flat and level surface to perform calibration
and let the scale adjust to room temperature.
3.
Press
to turn the scale on and wait for the scale
to stabilize.
4.
Press the
CAL
key. The display will show ‘CAL’.
5.
Make sure nothing is on the platform then press
CAL
to confirm the zero point.
6.
The display will then flash the required calibration
weight. (50g)
7.
Place the required
calibration weight
on the center of the
platform.
8.
The display will show
‘PASS’ then return
to normal weighing
mode. You may now
turn the scale off or
continue weighing.