Digital outputs
All outputs switch over to high resistance when the supply is smaller than 18V. When overload or
short circuit is detected the digital outputs are switched off after 100µs for a period of 2s. When time
is up the outputs get applied again.
Empty pipe output
Empty pipe
0V normal closed
High resistant
Filled, no flow
0V
0V normal open
0V
24V normal closed
High resistant
High resistant
24V
24V normal open
24V
High resistant
Pulse output
Empty pipe
0V normal closed
0V
Filled, no flow
0V
0V normal open
0V
24V normal closed
High resistance
0V
High resistance
24V normal open
High resistance
High resistance
Filled, flow
0V Pulses
0V Pulses
24V Pulses
24V Pulses
Upper limit output
Below lower limit
0V normal closed
High resistance
Between the limits
High resistance
0V normal open
0V
24V normal closed
High resistance
0V
High resistance
24V normal open
24V
0V
Above upper limit
0V
High resistance
24V
High resistance
Lower limit output
Below lower limit
0V normal closed
0V
Betwenn the limits
High resistance
0V normal open
High resistance
24V normal closed
24V
0V
High resistance
24V normal open
High resistance
24V
Above upper limit
High resistance
0V
High resistance
24V
19
PVDF-PE Plastic Ultrasonic Flow Meter
ULTRAFLO 5000