7.  Initial Inflate Method 

1. Wait until the power unit slowly inflates the mattress. However, initial inflation 

takes at least 30 minutes before the user is allowed to lie on the mattress.   

2. When air cells are completely deflated, it is recommended to inflate the mattress 

by connecting an external pump with the hose for quick inflation. When the air 
cells are inflated over 2/3, reconnect the hose to the power unit before the user is 
allowed to lie on the mattress. 

(The inflation tool maybe can be purchased from the dealer.) 

 

8.  Emergency CPR Deflate                      

 

A.

 

Emergency CPR Deflate: 

Pull open the CPR lid.

 (When the patient lies down, 

pull down hard on the plastic piece marked with CPR that drops from the chest 
position on the right side. Then, pull down the CPR lid for instant quick deflation.) 

 

B. 

Remember to resume the CPR lid for the Resume Function, 

otherwise it may lead to the danger of incomplete inflation. 

 

C.

 Pull open the CPR lid to deflate non-alternating air cells for storage . 

 
*Patent design No.# M347104    & ZL 200820136935.2 

 

 

9.  Power Unit Settings Contents   

 

 

Adjustment of the comfort range

     

   

The adjustment range is based on the user’s the bottom that can’t 
contact with the overlay.   
 

The softness and hardness of the air mattress can be selected from the 
number 1 to 10, to provide users comfortable.   
 

Suggestion: Normal setting in number 5 . 
Users weight limit: 30kg to 120kg. 

 

 

Summary of Contents for TS-105

Page 1: ...TRUE SOURCE User Manual Alternating Air Mattress Pump System Model TS 105 UM 105 E V1 0 Copyright 2010 by TRUE SOURCE Technology Co Ltd...

Page 2: ...4 4 Product Setup Contents 5 5 Product Features 5 6 Installation 6 7 Initial Inflate Method 6 8 Emergency CPR Deflate 7 9 Power Unit Settings Contents 7 10 Routine Maintenance and Cleaning 8 11 Stora...

Page 3: ...the absolute responsibility and obligation to guarantee its full effects Power Unit Do not open the power unit without guidance to prevent danger Do not insert items into any opening of the power uni...

Page 4: ...flicted with spinal cord injury bedridden for a long term not clearly conscious inflicted with bladder or bowel incontinence diabetes skin fragility physical weakness or malnutrition It is an extremel...

Page 5: ...streamlined system appearance and practicality 4 The cotton air filter is to be replaced outside Therefore the replacement can be done without opening the unit 5 The use of the double external fuse ho...

Page 6: ...e the air mattress evenly on the overlay The end with a connector to the power unit is the feet end Step 2 Check if the securing position of the air cells is centered Secure the overlay of the bed wit...

Page 7: ...en the CPR lid When the patient lies down pull down hard on the plastic piece marked with CPR that drops from the chest position on the right side Then pull down the CPR lid for instant quick deflatio...

Page 8: ...ivers must wear gloves and masks and protected with disinfection 2 Keep the coverlet even to prevent the formation of horizontal shear resulted from crinkles and the tightened skin 3 Do not wash the a...

Page 9: ...s off 2 2 Repair or replacement is needed if the lid is damaged 2 3 Check if the mattress connector is bent or falls off 2 3 Repair or replacement is needed if the connector is deformed and cannot be...

Page 10: ...ass I Type BF The Air Mattress Specification Allow different setup for marketing Standard State 5 Air cells 21pcs totally Material Coverlet PU membrane Elastic Fabric Air Cells Nylon PU Base Nylon PVC...

Page 11: ...Only function failure repair is covered during the warranty period The warranty does not include general maintenance e g replacement of the cotton filter cleaning setting adjustment or component renew...

Reviews: