7
7. Initial Inflate Method
1. Wait until the power unit slowly inflates the mattress. However, initial inflation
takes at least 30 minutes before the user is allowed to lie on the mattress.
2. When air cells are completely deflated, it is recommended to inflate the mattress
by connecting an external pump with the hose for quick inflation. When the air
cells are inflated over 2/3, reconnect the hose to the power unit before the user is
allowed to lie on the mattress.
(The inflation tool maybe can be purchased from the dealer.)
8. Emergency CPR Deflate
A.
Emergency CPR Deflate:
Pull open the CPR lid.
(When the patient lies down,
pull down hard on the plastic piece marked with CPR that drops from the chest
position on the right side. Then, pull down the CPR lid for instant quick deflation.)
B.
Remember to resume the CPR lid for the Resume Function,
otherwise it may lead to the danger of incomplete inflation.
C.
Pull open the CPR lid to deflate non-alternating air cells for storage .
*Patent design No.# M347104 & ZL 200820136935.2
9. Power Unit Settings Contents
Adjustment of the comfort range
:
The adjustment range is based on the user’s the bottom that can’t
contact with the overlay.
The softness and hardness of the air mattress can be selected from the
number 1 to 10, to provide users comfortable.
★
Suggestion: Normal setting in number 5 .
Users weight limit: 30kg to 120kg.