Measuring signal

4-20 mA

P+

Power supply

Transmitter

S1

P-

Sensor electronics

+

-

10...30 V

AC

DC

Power

Ammeter

or

4...20 mA input

Automaton

Supervisor

+

-

+

-

Red

Black

Transmitter

S2

+

-

Red

Black

Transmitter 

S1
[90°]

Scattered light

Transmitter

S2
[180°]

Direct light

 

 

                                    

    25 

   30

P

 

 

 

 

 

 

NOTE: 

The measuring output “4-20 mA" is a current drain, this means, the output channel has the same electrical behaviour as

a variable resistor. An active power supply for the measuring circuit is always needed. 

For connection to BAMOPHOX TUR, please see specific documentation msa436-01.

Signal in limit conditions: 

0 mA = wrong connected and / or dip-switches wrong adjusted
4 mA = 0 FNU (clearest water, pipe filled)
20 mA = 100% output acc. to adjusted range (pipe filled)
22 mA = over range
21 mA = general failure on electronics

When pipe is empty: 

If there are no water drops on the glasses the output will be a little bit higher than 4 mA (depending on the

refraction index of air-molecules). If there are water drops on the glasses the output is higher than 4 mA, but the signal depends on size
and quantity of water drops.

So it´s important to install the turbidity measurement system always in such a way that the pipe is filled at any time (ver-
tically or with a siphon). That’s valid not only for TURBICUBE but for all systems on the market.

5  WIRING

   

 

 

 

 

 

 

 

 

 

                                    

    25 

   30

-

 

 

 

 

 

 

 

 

 

 

Maximal load

[Ohms]

Voltage [V]

1200

1000

800

600

400

200

0

10                 15                20 

    25 

   30

P

 

 

 

Not permissible range

Permissible range

6  TECHNICAL FEATURES

Power supply:

24 V DC rated (10...30 V DC) 

Power consumption:

0,5 W

Output signal:

4-20 mA, for connection to the BAMOPHOX TUR 436

Ambient temperature

+5...+45 °C

Process temperature

+5...+100 °C

Material:

316L Stainless steel body / Terminal housing plastic PBT, IP 65

Accuracy:

±5% reading, ±1% full scale in use

Resolution:

0,001...0,2 FNU according to measuring range

Measuring ranges

Process connections 

(ND25...ND 100)

:

PN10 flanges 

(DIN 2633)

, Triclamp 

(DIN 11850)

L

CE Mark:

In accordance with low-voltage directive (2006/95/EG), EMC directives (2004/108/CEE)

ND

L: 

Flanges DIN 2633

L: 

Triclamp DIN 11850

ND 25 

274 

288

ND 32 

252 

266 

ND 40 

230 

244

ND 50 

190 

202

ND 65 

233 

249

ND 80 

208 

218

ND 100 

240 

260

ND 25...ND65 :

0.1...50 FNU

0.1...100 FNU

0.1...200 FNU

0.1...500 FNU

100...1000 FAU

ND 80...ND 100 :

0.1...50 FNU

0.1...100 FNU

0.1...200 FNU

0.1...300 FNU

100...1000 FAU