If a gauge is damaged, then follow the steps below:
✓
Locate the gauge and/or source.
✓
Do not touch or move the gauge.
✓
Immediately cordon off an area around the nuclear gauge and/or
source. A radius of 15 ft (5 m) will be sufficient. Do not leave
the area unattended.
✓
Keep all unauthorized personnel from the nuclear gauge.
✓
If a vehicle is involved, it must be stopped until the extent of
contamination, if any, can be established.
✓
The gauge user should perform a visual inspection of the
nuclear gauge to determine if the source housing and/or
shielding has been damaged.
✓
Use a survey meter to measure the dose rate at a distance of 3 ft
(1 m) from the gauge.
✓
Contact the company RSO (name and number given at the
beginning of this section). Provide the RSO with the following:
♦
The date, time, and location of the accident
♦
The gauge model and serial number
♦
The nature of the accident
♦
The location and condition of the gauge and/or source
♦
The dose rate at 3 ft (1 m) from the gauge
Appendix A–10