Figure 3–3. Standard Count Position
Press
〈
STD
〉
for the display:
-Standard Count-
xxx xxx
Take a new
Standard Count?
Press
〈
YES
〉
.
Place Gauge on
Spacer & both on
Block,Put Rod in
SAFE,Press ENTER
Check the gauge position. Press
〈
START/ENTER
〉
to start the
count.
Taking
Standard Count.
xxx seconds
remaining.
After the count is completed, accept the new count by pressing
〈
YES
〉
. The pass/fail tolerance is based on a 1% maximum allowed
variation for System 1 reference standard counts and a 1.2%
maximum allowed variation for System 2 standard counts.