multimeter BE47

Technical data

General characteristics

Parameter

Value

Diode test

Max. testing current of 1 mA, open-
circuit voltage of 2.8 V DC (typically)

Continuity test

Acoustic signal is emitted if the
resistance is less than 30 Ω

Battery testing current

9 V (6 mA); 1.5 V (100 mA)

Input impedance

>1 MΩ

V AC range

45 Hz to 450 Hz

A DC voltage drop

200 mV

LC display

3 ½ digits, 2000 count LCD, 0.5 inch
digits

Overload

OL will be displayed

Polarity

Automatic (no indication for positive);
minus (-) sign for negative

Measuring speed

2 x per second, nominal

Battery indicator

BAT

 is indicated if the battery voltage

drops below the operating voltage
threshold

Battery

 9 V monobloc battery

Fuses

mA, µA range: 0.5 A/600 V (fast
acting)
A range: 10 A/600 V (fast acting)

Operating temperature

0 °C to 50 °C (32 °F to 122 °F)

Storage temperature

-20 °C to 60 °C (-4 °F to 140 °F)

Relative humidity

Operation: <70 %
Storage: <80 %

Operating height above
sea level

Maximum 2000 m (7000 ft)

Weight

260 g (9.17 oz.)

Dimensions

121.5 x 60.6 x 40 mm 
(4.78" x 2.38" x 1.57")

Safety

This measuring device is designed for
indoor use and complies with
overvoltage category CAT II 600 V /
CAT III 300 V, pollution level 2.

Measuring ranges

Function

Measur-
ing range

Resolution

Accuracy

DC voltage
(V DC)

200 mV

0.1 mV

± (0.5 % +
2 digits)

2000 mV 1 mV

20 V

0.01 V

200 V

0.1 V

± (1.0 % +
2 digits)

600 V

1 V

AC voltage
(V AC, 50 / 60 Hz)

200 V

0.1 V

± (1.2 % +
2 digits)

600 V

1 V

Direct current
(A DC)

2000 µA

1 µA

± (1.0 % +
2 digits)

20 mA

10 µA

200 mA

100 µA

± (1.2 % +
2 digits)

10 A

10 mA

± (2.0 % +
2 digits)

Resistance
(Ω)

200 Ω

0.1 Ω

± (0.8 % +
2 digits)

2000 Ω

1 Ω

20 kΩ

0.01 kΩ

200 kΩ

0.1 kΩ

2000 kΩ

1 kΩ

± (1.0 % +
2 digits)

Battery test

9 V

10 mV

± (1.0 % +
2 digits)

1.5 V

1.0 mV

Note:
The accuracy is based on an ambient temperature of 18 °C to
28 °C and a relative humidity of less than 75 %.
The accuracy specification consists of two values:

% value referring to read value: Corresponds to the
accuracy of the structure to be measured.

+ digits: Corresponds to the accuracy referring to the
analogue to digital converter.

Scope of delivery

1 x Multimeter

1 x Safety measuring lines with test probes

1 x 9 V monobloc battery

1 x Getting started guide

Summary of Contents for BE47

Page 1: ...TRT BA BE47 TC2015 22 001 EN BE47 EN OPERATING MANUAL MULTIMETER...

Page 2: ...he device Always store the manual in the immediate vicinity of the device or its site of use Warning Read all safety warnings and all instructions Failure to follow the warnings and instructions may r...

Page 3: ...vice when the battery compartment or the housing are open Hazardous electric current Work on the electrical components must only be carried out by an electrically skilled person or an authorised speci...

Page 4: ...on 1 2 3 6 7 8 9 10 4 5 No Designation 1 Protective cover 2 Red measuring probe 3 Black measuring probe 4 Fuses 5 Battery compartment 6 V mA connection socket 7 COM socket 8 10 A socket 9 Rotary switc...

Page 5: ...Weight 260 g 9 17 oz Dimensions 121 5 x 60 6 x 40 mm 4 78 x 2 38 x 1 57 Safety This measuring device is designed for indoor use and complies with overvoltage category CAT II 600 V CAT III 300 V pollut...

Page 6: ...before opening the battery compartment Note Make sure that the surface of the device is dry and the device is switched off 1 Remove the protective cover 1 from the device 2 Loosen the screw at the bat...

Page 7: ...the red measuring line into the V mA measuring socket 3 Connect both measuring tips to the measuring object with correct polarity black to minus red to plus If the input voltage is negative a minus w...

Page 8: ...t resistance continuity or diode measurements switch off the current of the electric circuit and discharge all capacitors 1 Select the largest resistance measuring range using the rotary switch 2 Inse...

Page 9: ...been checked for proper functioning several times during production If malfunctions occur nonetheless check the device according to the following list Display segments are only faintly visible or flic...

Page 10: ...Trotec GmbH Co KG Grebbener Str 7 D 52525 Heinsberg 49 2452 962 400 49 2452 962 200 info trotec com www trotec com...

Reviews: