8. HARDWARE INSTALLATION
Refer to section 4 of this user manual before undertaking the installation of the Air XD instrument.
The installation location of the Air XD instrument is the responsibility of the installer and care should
be taken to ensure an appropriate position has been selected. Consider the location of a suitable
power supply and external fuses, access to a communications network and the protection of
cabling from damage.
8.1 Safety precautions
1. Secure the Air XD to a suitable mounting surface using the integrated
mounting brackets.
2. Ensure that the Air XD is mounted in an upright position.
3. Unlock and open the enclosure door.
4. Ensure that power is isolated before making electrical connections to the instrument.
5. Power supply voltage and frequency must match the instrument (refer to rating plate).
6. Ensure external switches or fuses are installed where applicable.
7. Run the required cables through the cable glands provided at the base of the enclosure.
8. Wire the cables into the relevant terminals as indicated on the internal plate (refer to
section 8.4).
9. Tighten the cable gland against the cable to ensure an IP seal.
10. Close and lock the door after use to maintain the IP rating of the enclosure.
11. Ensure that the air entry and exit ports are not restricted or covered.
Note:
When the door is open, care must be taken to ensure minimal internal contamination during
installation.