57 P5628.1600 Rev A
13
– PM D Av 2
14
– TSP Av 2
14
Alarm 1 threshold
Uint16_t
(As µg/m³)
15
Alarm 2 threshold
Uint16_t
(As µg/m³)
16
4-20mA Ch 1 assignment
Uint16_t
(0 = PM A
3 = PM D)
17
4-20mA Ch 2 assignment
Uint16_t
(0 = PM A
3 = PM D)
18
4-20mA Ch 1 scaling
Uint16_t
19
4-20mA Ch 2 scaling
Uint16_t
20
Average Ch 1 time
Uint16_t
(As minutes)
21
Average Ch 2 time
Uint16_t
(As minutes)
22
MODBUS baud rate
Uint16_t
0 = 4800
1 = 9600
2 = 14400
3 = 19200
4 = 38400
5 = 57600
6 = 115200
23
MODBUS Address
Uint8_t
Instrument Self-Test
On initial power on, the Air XD is programmed to perform a set of initialisation tests which are listed
and described below. Throughout general function, the Air XD will periodically perform these tests to
ensure correct operation.
•
OPC sensor comms check
Ensures communications and correct functionality of the OPC sensor.
•
OPC temp sensor test
Ensures the OPC sensor is operating within the specified safe temperature limits.
•
4-20mA self-test
During instrument start up and initialisation, each 4-20mA output will perform a calibration routine which
involves sweeping the output from 0-20mA before returning to 4mA. After this has been completed, the
output will return to configured functionality.
•
Data logging comms test
Ensures communications and correct functionality of the Air XD memory module.
•
EEPROM memory test
Ensures communications and functionality of the EEPROM is correct and that custom defined user
settings are not lost.