18
8.4 I/O terminals
Power, RS485, 4-20mA and relay connection terminal data is highlighted below.
Actuation type
Operating tool
Solid/stranded conductor
0.08 - 2.5mm² / 28 - 12 AWG
Conductor with ferrule
0.25 - 1.5mm²
Strip length
5 - 6mm / 0.2 - 0.24 inch
8.5 Power connections
For instruments connected to an AC power supply, it is the responsibility of the installer to ensure
that the instrument is installed with the following protection:
An external fuse or circuit breaker at a maximum of 5A.
Externally earthed.
See label below for details.
For instruments connected to a DC power supply, it is the responsibility of the installer to
ensure that the instrument is installed with a DC supply, meeting re-enforced insulation
requirements of EN61010-1 or equivalent.
For instruments connected to an AC power supply, ensure that the equipment is protected by
an external fuse or circuit breaker at a maximum of 5A.