苏
州
特
鲁
利
电
子
材
料
有
限
公
司
Suzhou Trojan Industry Material Co.
,
Ltd.
地地址:
江苏省苏州市工业园区新泽路 53 号 A 幢
电话:
0512-67580617
传真:
0512-67508229
Specimen holder:
The specimen holder is attached to the head drive shaft via two hex screws that tight on
the two flat sides of the shaft.
The specimen holder needs to be adjusted so that the pistons are approximately centered
in the holder holes. This adjustment is done by loosening the flange between the holder
and shaft coupler and rotating the holder into the correct position. Retighten screws and
nuts.
2.2 Height Adjustment for the Head /Specimen Holder
The specimen holder height needs to be adjusted to prevent the pistons from being ground during the operation of
the machine.
Step 1: Connect air and lift up head.
Step 2: Place the working wheel with the highest profile abrasive/polishing pad on the wheel. Using a single
wall box as a spacer, place the cardboard on top of the working wheel/abrasive combination.
Step 1: Rotate lever to
engage piston to lift head
Step 2: Lower head onto a
cardboard spacer