7
OPERATING INSTRUCTIONS
Your UV System is equipped with the ATS-281-2 UV Lamp Detector and Timer circuit board which is
designed to provide a continuous monitoring system of the UV lamp operational status. This is also
equipped with an internal fused surge protection system. This solid state electronic circuit board
provides the latest in UV lamp monitoring technology.
UV LAMP / CONTROL BOX START UP PROCEDURE - (4 Step Process)
1. Make sure the UV lamp is inserted per instructions into the quartz sleeve and UV chamber. Then
plug the UV lamp into the lamp connector.
2. Plug the electrical control box into a 120volt wall plug outlet.
3. Upon start up the UV lamp Detector and Timer circuit board will perform a self diagnostic test.
When power is applied the UV lamp will go on and a circuit board output check is performed as follows:
Quick Beep
Software version is displayed
The number 12 is displayed representing 12 months remaining
Green LED on, blinking decimal point to the right = System Check OK, UV System is operational.
4. Set the 1 Year Lamp Timer – WITH NEW UV LAMP
Make sure the control box is plugged into the wall outlet and the UV lamp is on. Press and hold the reset
button. Hold the reset button down for approximately 8 seconds until you see "re", continue holding
until you see the number 12 and hear it beep. At this point release the button. The 1 year lamp timer has
just been reset.
UV lamp Detector and Timer circuit board Lamp Operation and Alarm features:
MONTH LED
(Green)- Number of months remaining before the lamp has to be replaced.
DAY LED
(Orange)- Number of days remaining before the lamp has to be replaced.
ALARM LED
(Red)- Indicates a problem with the UV Lamp (lamp may be out or beyond its
rated life). Blinking decimal point indicates a properly functioning unit. Negative number indicates
the lamp is beyond its rated life.
LO
- Indicates the lamp is out and must be replaced. Beeping Alarm- Indicates problem or lamp life
will expire soon. Continuous Alarm- Lamp has expired or is off
SILENCE ALARM
- Press the reset button. There are a limited number of snoozes allowed.
NEW LAMP RESET-
Press and hold RESET button after 8 seconds "re" will appear, continue
holding until 12 appears on the display.