TRIRIDE s.r.l.
ManUso_TRIBIKE_eng_r01
01/06/2020
Pag. 45 di 49
USER MANUAL AND MAINTENANCE
17. EMC TABLES (TRIBIKE E - E PLUS)
Guide and manufacturer’s declaration – electromagnetic emissions
Tribike is designed to function in the electromagnetic environment indicated below. The customer or
user of the appliance must guarantee that it is used in such an environment
Emissions test
Compliance
Electromagnetic environment
RF CISPR 11 Emissions
Group 1
TRIBIKE uses RF energy only for internal
functions. Thus, RF emissions are very
low and therefore probably do not cause
any interference with nearby electronic
appliances
CISPR 11 Emissions
Class B
TRIBIKE is suitable for use in all those
environments,
including
domestic
environments and those connected
directly with a public grid low-voltage
power source that supplies buildings for
domestic use.
Harmonic emissions IEC 61000-3-2
Not applicable
Voltage/flicker fluctuation emissions
IEC 61000-3-3
Not applicable
Guide and manufacturer’s declaration – electromagnetic immunity
Tribike is designed to function in the electromagnetic environment indicated below. The customer or
user of Tribike must guarantee that it is used in such an environment.
Immunity test
IEC 60601 test level
Compliance level
Electromagnetic environment
Electrostatic
discharge (ESD) IEC
61000-4-2
on c- 6kV
on c- 2;4;6kV Floors must be wood, concrete or
ceramic tiles. If floors are coated
in synthetic materials, the relative
humidity must be at least 30%.
in air +- 8kV
in air +- 2;4;6;8kV
High-frequency
magnetic field
(50/60Hz) IEC
61000-4-8
3 A/m
30 A/m
Grid-frequency
magnetic
fields
should be at average levels of a
typical area in a commercial or
hospital environment.
Summary of Contents for 1776590
Page 1: ...EN MANUAL HYBRID PLUS MANUAL HYBRID PLUS USER MANUAL AND MAINTENANCE ...
Page 54: ...NOTES ...
Page 55: ...NOTES ...