2.
Place the thermocouple into the known, stable temperature
environment (such as an ice bath or dry well calibrator). Allow the
reading to stabilize.
3.
Use the arrow keys
▲ ▼
to scroll to the “T1
”
option and then press
ENTER
.
4.
Change the offset temperature reading (upper display) by
pressing
▼
until the desired temperature is reached.
5.
Press and hold the
SET
button for 3 seconds to return to normal
mode.
Replacing the Batteries
1.
Switch OFF the thermometer.
2.
Loosen the battery compartment screw and remove the compartment
door.
3.
Replace the AAA batteries.
4.
Replace the battery door and tighten the screw.