background image

 
 

Press and hold the 

SET

 button for 3 seconds to enter or exit the 

Setup mode. When the thermometer is in Setup mode, the 
display shows the 

SETUP

 display icon. 

 

 

 

Set Type J or K thermocouple 

 

1. 

Press 

 or 

to scroll to the “TYPE” setting (shown on the bottom left 

of the display). 

 

2. 

Press 

ENTER

 to choose the option. 

 

3. 

Press 

 to toggle between J and K. 

 

4. 

Press and hold the 

SET

 button for 3 seconds to save the setting and 

return to normal mode. 

 

Set Sleep/Non-sleep mode 

 

The thermometer enters sleep mode by default. The meter will 
automatically shut off after 20 minutes of inactivity (no buttons pressed). 
When the thermometer is in Setup mode, the display shows the SETUP 
display icon

 

1. 

Press 

 or 

to scroll to the 

SLP

 setting. 

 

2. 

Press 

ENTER

 to choose the option; 

On

 or 

OFF

 will appear in the 

display. 

 

3. 

Use the arrow key 

 to select ON or OFF. 

 

4. 

Press 

ENTER

 to store the new setting. 

 

5. 

Press and hold the 

SET

 button for 3 seconds to return to normal 

mode. 

 

Set Offset 

 

Use the offset option in Setup (

T1

 menu item) to adjust the 

thermometer’s readings to compensate for known temperature offset at 
a specific temperature. The adjustment range is ± 5.0°C or ± 9.0 °F. 

 

1. 

Plug the thermocouple into the input connector. 

 

Summary of Contents for TMP50

Page 1: ...ns on your purchase of the Triplett TMP50 Type K J Thermometer This meter This single input contact thermometer offers fast response and laboratory accuracy and works with K type or J type thermocoupl...

Page 2: ...ttery immediately when the low battery symbol appears on the LCD Do not touch the meter s circuit board for any reason static electricity or contamination could damage the sensitive components This in...

Page 3: ...isplay 11 Thermocouple Type Icon 12 Low Battery Indicator Icon Operation Connecting thermocouples 1 This meter accepts a K J type thermocouple with spade lugs sub miniature type with one spade wider t...

Page 4: ...hrs mins mins secs 1 Press MAX MIN to enter this mode 2 Press MAX MIN to step through maximum minimum and average readings 3 Press and hold MAX MIN for 3 seconds to return to normal mode Data Hold Pre...

Page 5: ...default The meter will automatically shut off after 20 minutes of inactivity no buttons pressed When the thermometer is in Setup mode the display shows the SETUP display icon 1 Press or to scroll to...

Page 6: ...t door 3 Replace the AAA batteries 4 Replace the battery door and tighten the screw Never dispose of used batteries or rechargeable batteries in household waste As consumers users are legally required...

Page 7: ...sions 220 x 63 x 28mm 8 7 x 2 5 x 1 1 Weight 200g 7oz Warranty Information Triplett Jewell Instruments extends the following warranty to the original purchaser of these goods for use Triplett warrants...

Reviews: