background image

 
 

 

Manual Measurement 

1. Pull and hold the trigger after aiming at the target. The SCAN icon 
will flash indicating that the target object temperature is being 
measured. The measurement result will be updated on the LCD. 
2. Release the trigger, the SCAN icon disappears, and the HOLD icon 
appears, indicating that the measurement has been stopped and the 
last measured value is held. 
 

 

Lock Measurement 

1. In the HOLD interface, press the SET button for 3s to enter the lock 
measurement setting interface, and turn on/off the lock 

measurement by pressing the 

 or 

 button. When the lock 

measurement is turned on, short press the LOG button to perform 
the time setting “00:00” for the lock measurement. At this time, the 
selected time position flashes, and the time value can be adjusted by 

pressing the 

 or 

 button. Set the time to "00:00" to turn off 

timing function. 
 
 
 
 
 
 
 
 
 
 

Summary of Contents for IRT350

Page 1: ...rchase of the Triplett IRT350 12 1 Circular Laser Infrared Thermometer This thermometer can quickly and accurately determine the surface temperature by measuring the infrared energy radiated from the...

Page 2: ...ing the thermometer Do not use the thermometer if it appears damaged Look for cracks or missing plastic Please refer to the emissivity information for the actual temperature Highly reflective objects...

Page 3: ...eading whichever is greater 32 F 3 0 F 0 1 F F 32 F 3 0 F or 1 5 of reading whichever is greater Temperature coefficient 0 1 C C or 0 1 C whichever is greater 0 1 F F or 0 1 F whichever is greater Rep...

Page 4: ...IN AV G DIF mode Lock measurement Data storage 99 sets Scheduled measurement Interval from 1 minute to 4 days up to 99 times Laser Circular laser wavelength 630nm 670nm output power 1mW class 2 laser...

Page 5: ...certification EN61326 1 2013 Laser safety standard EN60825 1 2014 Reference Standard JJG 856 2015 Features Circular laser indication which can indicate the target area under test more accurately and i...

Page 6: ...Scheduled measurement for occasions where timing temperature monitoCircular is required Tripod mount External Structure...

Page 7: ...tion Lock measurement indicator Temperature measurement alarm indicator Buzzer indicator Laser indicator Temperature hold indicator Temperature measurement indicator Low battery indicator Temperature...

Page 8: ...Operating Instructions Viewing the Last Measured Value In the off state short press less than 0 5s the trigger to turn on the thermometer and the measurement data held before last shutdown will be di...

Page 9: ...has been stopped and the last measured value is held Lock Measurement 1 In the HOLD interface press the SET button for 3s to enter the lock measurement setting interface and turn on off the lock meas...

Page 10: ...is reached the thermometer will automatically power off and store the last measured value Short press less than 0 5s the trigger to turn on the thermometer to view the measured value NOTE The measure...

Page 11: ...lect the storage location from 01 99 by pressing the or button If the selected location has stored data the temperature value and storage time will be displayed if there is no data will be displayed A...

Page 12: ...xit the measurement mode with data storage function In the measurement mode with data storage function press the LOG button for 3s until the screen starts to flash to exit Scheduled Measurement 1 In t...

Page 13: ...tion flashes and the value can be adjusted by pressing the or button NOTE The start time cannot be set less than the current system time otherwise the scheduled measurement will not be executed b Afte...

Page 14: ...D interface press the LOG button for 3s to enter the scheduled measurement log value query mode The screen will display the Auto Interval icon the LOG icon and the log group number In this mode press...

Page 15: ...ress and add or subtract 10 per second by long press Short press the SET button or pull the trigger to exit the system time setting NOTE System time needs to be reset after battery replacement or powe...

Page 16: ...tion is turned on and the measured temperature value is lower than the set low alarm limit the blue LED and LO indicator flash If the audible alarm function has been turned on the buzzer will beep Whe...

Page 17: ...value closest to the high low alarm limit and adjust it by pressing the or button Add or subtract 1 each time by short press and add or subtract 10 per second by long press 2 Emissivity Setting In the...

Page 18: ...HOLD interface short press the SET button until the audible alarm setting interface is displayed and turn on off the audible alarm by pressing the or button 5 Laser Indication Function Setting In the...

Page 19: ...easured area becomes larger The relationship between the distance and the spot size is as shown below D S 12 1 Field of View Make sure that the measured target is larger than the spot size The smaller...

Page 20: ...ace to be tested with masking tape or matt black paint with a high emissivity setting if it is possible wait for a period of time and measure the temperature of the tape or black paint surface when it...

Page 21: ...4 A3003 Alloy Oxidization Rough 0 3 0 1 0 3 Brass Burnishing Oxidization 0 3 0 5 Copper Oxidization Electric Terminal Board 0 4 0 8 0 6 Hastelloy Alloy 0 3 0 8 Inconel Oxidization Sand Blasting Elect...

Page 22: ...0 2 0 3 Iron Forging Passivation 0 9 Lead Rough Oxidization 0 4 0 2 0 6 Molybdenum Oxidization 0 2 0 6 Nickel Oxidization 0 2 0 5 Platinum Black 0 9 Steel Cold Rolling Steel Plate Rubbing Steel Plate...

Page 23: ...bide 0 8 0 9 0 7 0 8 0 9 Ceramics 0 95 Clay 0 95 Concrete 0 95 Cloth 0 9 Glass Convex Glass Smooth Glass Lead Boron Glass 0 76 0 8 0 92 0 94 0 78 0 82 Plates 0 96 Plaster 0 8 0 95 Ice 0 98 Limestone 0...

Page 24: ...wipe lens surface Use wet sponge or soft cloth to clean product exterior Do not rinse the thermometer or immerse it in water Battery Replacement Install or replace a 9V alkaline battery 1604A as foll...

Page 25: ...from the date of purchase This warranty does not apply to any of our products which have been repaired or altered by unauthorized persons in any way or purchased from unauthorized distributors so as i...