PON optical power meter - offset(calibration) setting
The followings will cause the test result differenct:
1.Different standard
2.Not use standard accessories.(Lab level 0 dB adapter, 0 dB
patch cord)
3.Dirty of connector
4.More than 1 year no calibration
User can do easy calibration as the following steps:
Press the " " key to the calibration menu as the picture.
1.Press the " " key to move the cursor.
2.Press the
key to increase the value
3.Press the
key to decrease the value
4.Press again the
key to save and exit.
Please note the adjustable range of each wavelength isfrom -5.0 to
+5.0, if the difference is out of the range, we suggest user send back
to manufacturer to do completely calibration.