4.20 Cleaning and Disinfection
Clean the Tremipen® only as advised.
Improper cleaning of the Tremipen® may
cause damages to the medical device or
possible infections
For cleaning, use only a lint-free cloth, which is free of
dust and use only the recommended
cleaning/disinfection agents:
23
The Tremipen®, being a medical device, must
be inspected/reevaluated every 2 years after the
first activation. If this inspection/reevaluation is
not done, NO warranty and liability concerning to
functionality and correctness of measurements of
the Tremipen® is guaranteed by Tremitas GmbH.
Inspection/reevaluation of the device is carried out by
Tremitas GmbH – for further information visit our
website www.tremitas.com or contact your local
Tremipen® distributor.
4.19 Inspection | Reevaluation
•
Keep the Tremipen® away from dust, fuzz, pets,
vermins and children to prevent staining.
•
Use non-aggressive surfactant based cleaning
agents (e.g. soap or detergents). Do not use bleach
or similar substances
•
Medicinal alcohol such as Isopropanol (max. 70%)
•
Do NOT use aggressive cleaning agents, brushes
with metal or nylon bristles or sharp or metallic
cleaning articles, such as knifes, hard spattles etc.
They can damage the surface.
•
Avoid that the Tremipen® comes into contact with
running water.
•
Do NOT put the Tremipen® into the dishwasher.
•
Wipe-off the Tremipen® with a slightly moistened
cloth. Wipe it dry afterwards.
•
For disinfection, take a cloth and moisten it a little
with one of the recommended agents. Wipe the
Tremipen® immediately dry disinfection, to prevent
liquids from entering the Tremipen®.