Electromagnetic environments
Electromagnetic environment
Treatlite uses RF energy only for its internal function. Therefore, its RF emissions are very low and are not likely
to cause any interference in nearby electronic equipment. Treatlite is suitable for use in all establishments
including domestic establishments and those directly connected to the public lowvoltage power supply network
that supplies buildings used for domestic purposes. For more information about electromagnetic environment
see below.
Electromagnetic emissions
Treatlite is intended for use in the electromagnetic environment specified below. The customer or the user of
Treatlite should assure that it is used in such an environment.
Electromagnetic immunity
Treatlite is intended for use in the electromagnetic environment specified below. The customer or the user of
Treatlite should assure that it is used in such an environment.
Emissions test
Compliance
Electromagnetic environment - guidance
RF emissions
CISPR 11
Group 1
Treatlite uses RF energy only for its internal function. Therefore,
its RF emissions are very low and are not likely to cause any
interference in nearby electronic equipment.
RF emissions
CISPR 11
Class B
Treatlite is suitable for use in all establishments including
domestic establishments and those directly connected to the
public lowvoltage power supply network that supplies buildings
used for domestic purposes.
Harmonic emissions
IEC 61000-3-2
Class A
Voltage fluctuations/ flicker
emissions
IEC 61000-3-3
Complies
Immunity test
IEC 60601 test level Compliance level
Electromagnetic environment – guidance
Electrostatic discharge
(ESD)
IEC 61000-4-2
+/– 6 kV contact
+/– 8 kV air
+/– 6 kV contact
+/– 8 kV air
Floors should be wood, concrete or
ceramic tile. If floors are covered with
synthetic material, the relative humidity
should be at least 30%.
Electrical fast transient / Burst
IEC 61000-4-4
N/A
Surge
IEC 61000-4-5
N/A
Voltage dips, short interruptions and volta-
ge variations on power supply input lines
IEC 61000-4-11
N/A
Power frequency (50/60 Hz) magnetic
field
IEC 61000-4-8
3 A/m
3 A/m
Power frequency magnetic fields should
be at levels characteristic of a typical
location in a typical commercial or hospital
environment.
NOTE: UT is the a.c. mains voltage prior to application of the test level.