TROUBLESHOOTING
SPECIFICATIONS
Power supply
2*AAA alkaline batteries
Display moder
Digital LCD V.A.36x41mm
Measurement mode
Oscillographic testing mode
Measurement range
Accuracy
Normal working condition
Storage & transportation
condition
Measurement perimeter
of the wrist
About 13.5cm-19.5cm
Weight
Approx.120g(Excluding the dry cells)
External dimensions
Attachment
Approx.80×65×22mm
2*AAA alkaline batteries,user manual
Mode of operation
Continuous operation
Degree of protection
Type B applied part
Protection against
ingress of water
IPXO
Software version
Device classification
V01
Internally Powered ME Equipment
WARNING: No modification of this equipment is allowed.
Pressure: 0kpa-40kpa
(
0mmHg-300mmHg
)
pulse value:(40-199)times/minute
Pressure:
15
ć
-25
ć
within±0.4kpa(3mmHg)
10
ć
-40
ć
(out of 15
ć
-25
ć
)
vithin±0.8kpa(6mmHg)
pulse value:±5%
Temperature:-20
ć
to 60
ć
Relative humidity:10% to 93%
Temperature:5
ć
to 40
ć
5HODWLYHKXPLGLW\
Atmospheric pressure: 86kPa to 106kPa
This section includes a list of error messages and frequently
asked questions for problems you may encounter with your blood
pressure monitor. If the products not operating as you think it
should, check here before arranging for servicing.
PROBLEM SYMPTOM
CHECK THIS
REMEDY
No power
Low
batteries
Error
massage
Display is dim or
will not light up.
Batteries are exhausted.
Replace with new batteries
Insert the batteries
correctly
Replace with new batteries
Batteries are inserted
incorrectly.
Show on
the display
Batteries are low.
E 1 shows
Data communication
has failed
Make sure that phone’s
Bluetooth is on or within
the distance range
E 2 shows
The cuff is very tight
Refasten the cuff and then
measure again.
E 3 shows
The pressure of the
cuff is excess.
Relax for a moment
and then measure again.
E 9 shows
Product has not been
activated.
Reactivated
E 10 or
E 11 shows
The monitor detected
motion while measuring.
movement can affect the
measurement.Relax for a
moment and then
measure again.
E 20 shows
E 21 shows
Relax for a moment and
then measure again.
EExx,shows on
the display.
A calibration error
occurred.
Retake the measurement.
If the problem persists,
contact the retailer or our
customer service
department for further
assistance.Refer to the
warranty for contact
information and return
instructions.
The measurement
process does not detect
the pulse signal.
Loosen the clothing on
the arm and then
measure again.
Measure incorrectly.