Data transmitting

Data transmitting to 

wireless wellness system:

-If SUCCEED, the symbol 

disappears;

-If FAIL, the symbol 

remains.

Data Transmission

Troubleshooting

Data Transmission

Error Prompt

Error

Description

Solution

Overload. The device will 

power off.

Stop using this scale for 

measurement.

Low Battery. The device 

will power off in three 

seconds.

Replace all three batteries at 

the same time. Please purchase 

the authorized batteries for 

replacement.

Please check below items:

-Bluetooth is ON.

-App Collector is ON.

-Both devices are within the 

transmission distance of Bluetooth.

Failure of pairing up .

When Measuring ...

Problem

Root Cause

Solution

Incorrect posture

Abnormal 

measuring results:

- Too high; OR

- Too low; OR

- Huge difference 

between two recent 

measurement.

Please step on the platform 

barefooted and stand still.

The device is located on 

the soft ground such as a 

carpet OR on a rugged 

surface.

Please place the device on a 

flat, hard surface.

CAUTION

Interference may occur in the vicinity of equipment marked with the following symbol        . And 

the Analyzer may interfere vicinity electrical equipment.
To enable the data transmission function, this product should be paired to a Bluetooth 4.0 end. 

How to mitigate possible interference?

1. The range between the Analyzer and the Bluetooth end should be reasonably close, from 1 

meter to 10 meters. Please ensure no obstacles between the Analyzer and the Bluetooth end 

so as to obtain quality connection.

2. To avoid interference, other electronic devices (particularly those with Bluetooth transmission 

/ Transmitter) should be kept at least 1 meter away from the Analyzer. 

With GBF-1144-B successfully pair-up with your 

smartphone, the measurement data will be 

transmitted to your mobile via Bluetooth.

The symbol      will disappear after successful data 

transmission, and you may check your personal 

health data stored in your smartphone.

If the data transmission fails, the symbol      will 

remain. The pending measurement data will be 

temporarily kept in the scale and transmitted to your 

smartphone when next measurement is complete.

There has data doesn’t 

upload. (The notebook

icon will be displayed.) 

Upload the data.

Summary of Contents for GBF-1144-B

Page 1: ...safe usage of this product and keep the manual well for further reference in case you have problems Thank you very much for selecting the Transtek Body Fat Analyzer GBF 1144 B TEL 86 760 88282982 www...

Page 2: ...ert the Batteries Install App and Pair up Start Measuring Select Measurement Unit Daily Measurement Data Transmission Data Transmission Troubleshooting Error Prompt When Measuring When Data Transmitti...

Page 3: ...MMUNITY of the device Be aware that misuse of electrical equipments can cause electric shock burns fire and other hazards Warning that the body fat analyzer should not be used adjacent to or stacked w...

Page 4: ...tteries 1 5V per each 3 User Manual Overview Device Components LCD Display Battery Compartment UNIT Button LCD Display Overview Pending to transmit to wireless wellness system Stone Data Transmitting...

Page 5: ...period of time it is advisable to remove the batteries before storing Transtek Body Fat Analyzer GBF 1144 B applies BIA Bio impedance Analysis technology A small amount of weak current flows through t...

Page 6: ...ted Stand still and keep full contact with the electrodes until the LCD displays the weight and stops blinking the unit When the weight is locked the symbol BIA blinks indicating the Bioelectrical Imp...

Page 7: ...cur in the vicinity of equipment marked with the following symbol And the Analyzer may interfere vicinity electrical equipment To enable the data transmission function this product should be paired to...

Page 8: ...ment Batteries not yet installed Worn batteries Low battery Install the batteries Please refer to Insert the Batteries Replace all three batteries at the same time Please purchase the authorized batte...

Page 9: ...F applied part When carrying out usual maintenance please ensure practice of the following Do s and Don ts Maintenance Warranty DO use a dry soft cloth to wipe the dust DO use a wet soft cloth dipped...

Page 10: ...se undesired operation Changes or modifications not expressly approved by the party responsible for compliance could void the user s authority to operate the equipment FCC RF Exposure Guidelines Safet...

Page 11: ...dance and MANUFACTURER s declaration electromagnetic IMMUNITY for all ME EQUIPMENT and ME SYSTEMS IEC 60601 test level Compliance level Electrostatic discharge ESD IEC 61000 4 2 Electromagnetic enviro...

Page 12: ...d RF transmitters an electromagnetic site survey should be considered If the measured field strength in the location in which the device is used exceeds the applicable RF compliance level above the de...

Reviews: