4.30 PC-300–PC-550 Procedural/Fixed Height/Trauma/Surgical/ OB/GYN/Electric Stretcher Caster
Chapter 4: Removal, Replacement, and Adjustment Procedures
TranStar® PC-100/PC-200/PC-300/PC-350/PC-400/PC-450/
Page 4 - 97
PC-500/PC-550/TR-500/SG-500/OB-450/OB-550/EL-550
Series Stretcher Service Manual (man115re)
4
WARNING:
Use two people for this procedure. Failure to do so could result in
personal injury and equipment damage.
7. Lift the corner of the stretcher, and temporarily block it up with the
8" x 2" x 4" (203 mm x 51 mm x 102 mm) lumber.
Replacement
1. Align the caster with the color mark (G) on the base of the stretcher
(see figure 4-52 on page 4-97).
Figure 4-52. Caster Brake Orientation
2. Slide the caster into the base of the stretcher (see figure 4-51 on
3. Using the ratchet and 7/16" socket, install the caster screw (D), and use
blue Loctite® adhesive to ensure that the screw will stay in place.
NOTE:
Use the 13 mm socket to install the caster screw on the
surgical stretcher
only
.